16
Manual TECHKON DENS
2.3 Measurement function T: Density measurement
in transmission
T mode
measures the densities in transmission
.
Density measurement in T mode:
1. Select the T mode with the MODE button.
2. If necessary do a calibration without film by
pressing the CAL button.
3. Position the device on the measurement
patch selected.
4. Start the measurement with the DENS but-
ton.
The measurable density in transmission de-
pends on the luminosity of the light table. Av-
erage bright light tables allow you to measure
densities up to D = 4.00 and over.
Please note: The emulsion side of the film has
to be always on the upper side.
2.4 Measurement function T %: Dot % measurement
in transmission
T % mode
measures the screen densities in transmission.
Measurement of dot percentage values:
1. Select the T % mode with the MODE but-
ton.
2. If necessary do a calibration on film (0 %)
with the CAL button.
3. Position the device on the measurement
patch selected.
4. Start the measurement by pressing the
DENS button.
Please note: The emulsion side of the film has
to be always on the upper side.