TROUBLE-SHOOTING SCHEME
Vll:1. FUNCTIONALTROUBLESHOOTING SCHEME
Fault
A. The waterpump does not
start
B. The hotplates do not
heat up
C. Poor reflux of solvent
D. Solvent leakes out
F. Hotwater leakage
Probable Cause
1. Nocurrentto the pump
1. Too low temperature in
tank
2.
1.
2.
3.
4.
1.
2.
E. Too long evaporation time 1.
2.
3.
4.
1.
2.
3.
Heating valveclosed
Too low water flow in the
condensers
Too low temperature on
the hotplates
Extraction Cup contamina-
ted on bottom surface
Recovery Valve closed
Extraction Cup incorrectly
placed
Rubber Sleeve between
condenserand extraction
cup isdamaged
Too low temperature on
the hotplates
Extraction Cup contami-
nated on bottom surface
Poor ornoairf low
Sealing membrane dama-
ged
Unsatisfactory tightening
of theconnections
Rubber seal in the connec-
tion isdamaged
Shaft seal on the water
pump isdamaged
Corrective Measure
Connect the cable. Switch on
Mains
Screw the thermostat clockwise
(Check temperature with a ther-
mometer in the tank)
Open the Heating valve
Increase the water flow
Adjust thermostat in watertank
Clean the Extraction Cup
Open Recovery Valve. Turn
counter clockwise
Place Cups on center of the hot
plates
Replace the rubber sleeve
Adjust thermostate in water-
tank
Clean the Extraction Cup
Check that evaporation valve is
open and that the PVC-tubing
between Service and Extraction
Unit is not blocked. Check the
pressure pump according to
para Vl:3
Replace membrane according
topara Vl:7
Fasten the connections
Replace the seal
Replace the shaft seal accord-
ing topara Vl:2
Vll:2. METHOD TROUBLE SHOOTING SCHEME
Fault
A.
Too great deviation bet-
ween results from tests on
the same sample
B.
Solvent (with fat) frothing
up on the thimble during
rinsing or evaporation
Probable Cause
1. Too coarse structure of
the sample
2. Unsatisfactory mixing of
sample prior to weighing
3. Unsatisfactory weighing in
of sample or cups
4. Too short extraction time
5. Poor reflux
1. Too high fat content
2. Too high temperature
on the hot plate
Corrective Measure
Grind or homogenize to a finer
structure
Check the mixing procedure
Check the weighing procedure
Check the extraction time
See section C under Vll:1
Weigh in less sample
Adjust the thermostat in the
watertank
19