5. Maintenance
& Cleaning
46
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
2021-07
5.5
Preventive Maintenance Plan for SUNRISE
This preventive maintenance plan is for standard throughput instruments. For
instruments that are used in high throughput, the maintenance intervals might be
shorter.
5.5.1
Daily
•
No daily maintenance is required.
5.5.2
Weekly
•
Clean the cover and the plate transport with a mild detergent.
CAUTION
NEVER USE ACETONE AS IT WILL DAMAGE THE COVERS.
5.5.3
Every Six Months
•
Clean the filters using an optical cleaning solution (lens tissue recommended)
5.5.4
Yearly (Customer or Service Technician)
•
Either done by customer:
•
Perform the QC Pac 2 Test (see QC Pac Instructions for Use)
•
Or done by service technician:
•
Perform extended operational quality check
5.5.5
Every Four Years (Service Technician Required)
•
Replace the lamp and the filters
•
Perform extended operational quality check