4. Error
Messages and Trouble Shooting
2021-07
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
39
Error Messages
Description
Trouble Shooting
Filter Errors
Already inserted
Filter already inserted
Check if the filter is properly inserted.
No filter carriage detected
The instrument does not
detect the filter carriage
Insert filter. If a filter has been already
inserted, check the filter carriage for dirt or
damage. Contact your local service
engineer if the error continues.
No measurement filter
defined
The measurement filter is not
defined
Define filter.
No reference filter defined
The reference filter is not
defined
Define filter.
Illegal filter carriage
position
Internal firmware or electrical
error
Check the filter carriage for dirt or
damage. Contact your local service
engineer if the error continues.
Wavelength … nm not
available
The defined reference or
measurement filter is not
available on the inserted filter
carriage
Change filter carriage or check filter
values for incorrect input.
Filter carriage not defined,
Type.. Number…
Wrong, damaged or not
defined filter carriage
inserted
Check the filter carriage to see if it is
correct or check the filter carriage for dirt
and damage.
ADC Electronic Error
Offset 340 not adjusted
Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the Setup
program. Contact your local service
engineer if the error continues.
Offset 400 not adjusted
Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the Setup
program. Contact your local service
engineer if the error continues.
No wavelength defined
Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the Setup
program. Contact your local service
engineer if the error continues.
Area 400 not adjusted
Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the Setup
program. Contact your local service
engineer, if the error continues.
Area 340 not adjusted
Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the Setup
program. Contact your local service
engineer if the error continues.
E2Pot Overflow
Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the Setup
program. Contact your local service
engineer if the error continues.