12
The settings of the following items can be changed
from the Home Screen.
Sampling series (SERIES) (page 25)
Sampling frequency (SAMPLE) (page 25)
Analog-digital conversion bit depth (AD) (page 25)
Monitor channel (MONITOR) (page 25)
5-7. Recording and playing data
Recording data can be saved on the LX-1000 SD card.
Install AR-LXAO1000 analog output amplifiers into the
LX-1000 as necessary.
ATTENTION
Data saved to an LX-1000 SD card can be played back.
Data saved to a computer cannot be played back.
1
Turn the LX-1000 on.
Wait until the Home Screen appears on the display.
2
Set the measurement conditions or load them.
Press the MENU/HOME button to open the Menu
Screen and make necessary settings (page 20).
To load measurement conditions saved on the SD
card, select PARAMETER on the MISC screen. Then,
select LOAD. To save the current setting values,
select SAVE (page 33).
ATTENTION
If an input/output amplifier module is changed, the
measurement conditions will be reset.
3
Record or play data.
Recording
Press the
0
REC button to make the system record
ready.
If a start trigger has been set, recording will start
when a trigger condition is met.
If no start trigger has been set, press the
7
FWD
button to start recording (page 16).
Playback
Press the
7
FWD button to start playback. This will
start playback of the most recently recorded data.
To play another file, press the MENU/HOME button,
and select FILE. Then, select OPEN FILE to open the
playback folder selection window and select a file
for playback (page 18).
5-8. Synchronization function
For details about master and slave unit connections,
refer to “Synchronization function” in the LX-1000
Instructions for Use.
See “8-3-5. SYNC SETTINGS” on page 31 for informa-
tion about master and slave device settings.
5-8-1. Synchronized recording settings
If you set SAMPLING FREQ., AD BITS and DEVICE (record-
ing destination) for the master unit, they will also be set
for slave units automatically. These settings cannot be
changed on the slave units.
The number of recording channels can be set sepa-
rately for the master and slave units. An error will occur
if recording is conducted at a transmission rate higher
than the slave units can record.
Confirm that the numbers of recording channels on the
slave units are suitable before starting recording.
o
Setting triggers on the slave units is not possible.
o
The following amount of time is necessary to start
synchronized recording from a stopped state.
Sampling frequency Starting delay
96 kHz – 12 kHz
About 2 seconds
6 kHz
About 3 seconds
3 kHz
About 4 seconds
1.5 kHz
About 5 seconds
o
Level triggers and external triggers become effective
ten seconds after the unit becomes record ready.
o
If recording cannot be continued with the master
unit because, for example, it does not have enough
recording media capacity, recording will stop at that
moment.
If recording cannot be continued with a slave unit
because, for example, it does not have enough
recording media capacity, recording will stop only
for that unit. Other units will continue recording, but
pausing will stop recording.
5. Basic operation
Summary of Contents for ER-LXRC1000
Page 1: ...Remote Control Unit Instructions for Use Z ER LXRC1000 D01349910F...
Page 37: ...37...
Page 38: ...38...