6. Recording
6-1. Order of procedures
Set recording conditions
c
Set recording destination
c
Calibration
c
Set triggers
c
Make other settings
c
Start recording
c
Stop recording
6-2. Setting recording conditions
On the Menu Screen, make settings for the sampling
frequency, analog-digital conversion bit depth, voice
memos and modules (page 25).
6-3. Setting recording destination
Set the file name for recording.
Set the recording destination on the FILE screen
(page 31).
o
Select FORMAT on the FILE screen to format the SD
card loaded in the LX-1000.
o
If the ID number exceeds 999, recording will stop.
6-4. Calibration
Set the value used to convert sensor output to physical
quantities.
6-4-1. Equivalent input calibration
In the physical quantity conversion fields, set the rated
output and rated capacity indicated in the sensor test
report (page 26).
6-4-2. TEDS calibration
TEDS data is not read automatically when the power is
turned on or when the TEDS sensor is changed.
TEDS data must be manually updated by pressing the
Update button (page 25).
6-4-3. Zero balancing
Zero balance can be adjusted for channels with strain
amplifiers that have the amplifier mode set to ST.
1
Press the Zero Bal. (F3) button on the Home
Screen.
2
Turn the VALUE dial to change the selec-
tion position, and press the ENTER button to
change channel selection for zero balancing.
3
Press the Execute (F3) button to conduct zero
balancing.
o
Press the Initialize (F2) button to cancel the zero
balancing adjustment results.
o
Press the Check All (F4) button to select or deselect
all channels.