27
7.2
Replace the main HEPA filter
Figure 20
CAUTION
Handle the filter with extreme care.
Once the filter is damaged its filtration capabilities are limited.
Make sure you don’t
damage or touch:
•
the filtration material;
•
the gasket on the filter frame.
1.
Turn the two knobs (Figure 20, pos. 5) on the main HEPA filter compartment panel
(Figure 20, pos. 4) counter clockwise until the panel is loose.
2.
Open the compartment panel.
NOTE
: There is a ground wire (Figure 20, pos. 3) connecting the compartment panel and the
main ICon Pro unit.
DO NOT
disconnect it.
3.
Pull up the two yellow locking levers (Figure 20, pos. 2) to release the filter.
4.
Carefully pull out the filter (Figure 20, pos. 1) from the compartment.
5.
Put the old filter in the plastic bag supplied with new filter.
6.
Seal the plastic bag and follow the instruction on the bag for disposal.
WARNING
•
Always switch off and unplug the unit before replacing the main HEPA filter.
•
Always use the replacement filters recommended by your ICon Pro supplier.