ATP Operator Instructions
|Selecting an Installation Site
13
Chapter 3 |
SELECTING AN INSTALLATION SITE
Overview
This unit is a precision instrument. To operate this unit properly and to obtain maximum performance,
install it in a location that meets the following conditions. The unit stores a certain amount of medications
in the tablet canisters while dispensing the required amount. Temperature and humidity conditions must
be closely monitored since medications are temporarily stored.
•
Install the unit indoors.
•
Select a location not exposed to direct sunlight and that is free from dust.
•
Install it on a sturdy, level floor.
•
Select a location with low humidity. But, avoid a location that is too dry.
•
Do not install it near a wash basin or kitchen sink.
•
Install it in a location without stoves, gas ranges and other heat-generating equipment.
•
Do not install it near equipment that generates strong magnetic fields.
•
Select a location away from TV or radio sets. Otherwise, the unit may distort the picture and sound of these
appliances.
•
Leave a space of at least 10 cm from the rear side of the unit to ensure that the internal fan is not blocked.
•
Ambient temperature:15 to 28
℃
•
Ambient humidity: 20.to 70 %R.H.
•
At an altitude up to 6,562 feet (2000 m)
•
Using a mains supply voltage that does not exceed 115VAC ± 10%, 60Hz. <6 amperes, with transient over
voltages II in accordance with IEC-60664-1
•
Consistent with Pollution Degree 2 in accordance with IEC-60664-1
•
Use a grounded AC wall outlet to prevent electric shock. Have authorized personnel install a grounded AC
wall outlet in a location that is not provided with one.