tbs electronics SR12 Operation And Service Manual Download Page 14

Page  14 

become clogged with dirt 

 

 
 
 
 
 
TECHNICAL DATA 

 

Maximum Capacity 

12 Liters (3.2 gal) 

19 Liters (5.0 gal) 

Type of Distillation 

Simple 

Simple 

Volume of Thermal Oil 

5 Liters (1.5 gal)

 

14 Liters (4.0 gal)

 

Voltage 115V/60Hz 

115V/60Hz 

Current 

9 Amps 

17.5 Amps 

Working Temperature 

40 - 200°C. (424°F.) 

40 - 200°C. (424°F.) 

Cycle Time 

5 -6 Hours 

5 -6 Hours 

Condenser 

Air Cooled 

Air Cooled 

Size (W x D x H) 

59 X 59 X 145 cm 

61 X 76 X 173 cm 

Weight 

74 kg 

135 kg 

 

 

 
LIST OF PARTS

 

 

 

1. Distilling Chamber 

2. Camber Cover 

3. Electrical Box Cover (Expl Prf) 

4. Electric Box Housing (Expl Prf) 

5. Bushing Carrier (Expl Prf) 

6. Fraction Switch Cable 

7. Electrical Heating Element(s) 

8. Thermal Oil Vessel 

9. Fan Motor Junction Box (Expl Prf) 

10. Distilling Condenser 

11. Cooling Fan Motor (Expl Proof) 

12. 3-Way Fraction Switch (Expl Prf) 

13. Heater Cable Sheath Pipe (Expl Prf) 

14. Solvent Receiver Tanks 

15. Solvent 

16. Oil Discharge 

17. Thermal Oil Level Indicator 

18. Working Thermostat T-1 

19. Timer 

20. Fraction Thermostat T-4 

21. Indicator Window 

22. Vapor Temperature Gauge 

23. Oil Bleeding Valve 

24. Locking Knob 

25. Cover Gasket 

26. Arm for Cover 

27. Vapor Collector Duct (Manifold) 

28. Solvent Outlet Nozzle- Left 

29. Solvent Outlet Nozzle- Right 

30. Oil Temperature Gauge 

31. Oil 

32. Bag Holder/Max Level of Solvent 

33. Main Power Electric Cable 

34. Motor Cable Sheath Pipe (Expl Prf) 

35. Unit Support 

36. Shell 

37. Heater On Indicator-Orange 

38. Main Power Indicator - Green 

39. Alarm Indicator Light - Red 

40. Timer Module 

41. Working Thermostat T-1 

42. Safety Thermostat T-2 

43. Safety Thermostat T-3 

44. Fraction Thermostat T-4 

45. Safety Relay 

46. Heater Power Relay 

47. Fraction Switch Junction Box 

 

Summary of Contents for SR12

Page 1: ...Triangle Biomedical Sciences Inc Solvent Recovery System Operation And Service Manual SR12 Version 1 2 Serial ___________________...

Page 2: ...ION 4 Application 4 Function 4 Separation Module 4 Safety Measurers 4 OPERATIONAL SAFETY 4 Location 4 Solvent Containment 4 Safe Exposure Temperature 5 OPERATING PROCEDURES 6 Purpose 6 Procedure for A...

Page 3: ...Page 3 GLOSSARY 15...

Page 4: ...ansfer the benefit of such warranties to the user This warranty supersedes and is given in lieu of all implied warranties and is void if user does not provide unit with continuous ample electrical pow...

Page 5: ...illation using the Separation Module and Baffle The Switching Module allows an operator to set a fraction temperature at which the distillate of the different solvents will be separated The fraction t...

Page 6: ...on the floor under your recovery unit This pan is recommended for use as a safety pan to prevent solvents from spilling onto the floors in case of an over flow due to operator error Appropriate 3 gal...

Page 7: ...taminated solvent in the distilling chamber up to 3 gallons NOTE Take special care NOT to allow any contaminated solvent to enter the Vapor Collector Duct Manifold when you are filling the distilling...

Page 8: ...hen operating This could create a dangerous situation Xylene 1 Place clean and dry collection containers under each of the lower distillate solvent outlet nozzles left and right 2 Verify that the Ther...

Page 9: ...overy cycle has been completed let the unit cool to at least 50 C prior to opening the secured top Premature opening will allow a release of solvent fumes into the surrounding atmosphere and can also...

Page 10: ...156 377 Cyclohexane 81 260 Cyclohexanon 155 419 1 2 DiChlorethane 84 412 DiEthyl Cellosolve 121 208 DiEthyl Ketone 104 445 Ethanol 82 400 Ethyl Acetate 79 427 Ethyl Amyl Keytone EAK 160 Ethylene Glyc...

Page 11: ...113 46 80 Carbon Tetrachloride Tetrachloromethane 78 110 1 1 2 2 Tetrachloroethane Acetylene Tetrachloride 147 160 Tetrachloroethylene Pestabil Vors Perchloroethylene Per 121 150 Trichloroethane 1 1...

Page 12: ...o control the temperature of the thermal oil that transfers heat to the distilling chamber Range 40 200 C Note The oil temperature is generally 15 20 C higher than the temperature of the contaminated...

Page 13: ...time lasts much longer than 4 hours 1 Contaminated solvent contains too much water 2 Thermal oil level is low 3 Thermal oil is old and dirty 4 Heating elements need cleaning 1 Replace the solvent 2 Ad...

Page 14: ...er 11 Cooling Fan Motor Expl Proof 12 3 Way Fraction Switch Expl Prf 13 Heater Cable Sheath Pipe Expl Prf 14 Solvent Receiver Tanks 15 Solvent 16 Oil Discharge 17 Thermal Oil Level Indicator 18 Workin...

Page 15: ...g Thermostat T1 384010 Safety Thermostat T2 465000 White Warning Light 465001 Red Warning Light 465002 Green Warning Light 475000 Ventilator Motor 468402 Heating Element 385007 Timer 425002 Control Kn...

Page 16: ...since the distlland mixture is forced into the area in which only pure material is desired Fraction A portion of the collected separated material recovered from the head of the still Fractional Disti...

Reviews: