Operating Instructions – TN50
34
10.3 RECOMMENDED MAINTENANCE AND SERVICE SCHEDULE
Daily
Remove frozen section waste from the cryostat chamber after every day
of usage.
Disinfection if necessary.
Weekly
Cleaning of blade holder (see section 10.3).
Cleaning of window.
Monthly
Check of temperature sensor in cryochamber.
Cleaning of cryostat ventilation (see section 10.4).
Yearly
Complete Service
(performed by authorized Tanner Scientific
service technician)
check of all functions
complete defrost
removing of microtome
oiling of movable parts
check of cooling system
check of driving system
complete disinfection / cleaning / drying
Summary of Contents for TN50
Page 1: ...OPERATING INSTRUCTIONS Cryostat TN50 INS2000GB 2018 10 003 00 ...
Page 2: ...Operating Instructions TN50 2 ...
Page 40: ......
Page 41: ......