Operating Instructions – TN50
18
7.5 List of available software settings
Chamber temperature
-35 °C to ambient
The recommended working temperature is -15 °C to -25 °C
Object cooling
ON / OFF
If the instrument is equipped with the optional
object cooling system, its function can be activated
or deactivated.
Object temperature
-50 °C to 0 °C
If the instrument is equipped with the optional
object cooling system, its function can be activated
or deactivated.
Quick-freeze
ON / OFF
If the instrument is equipped with the optional fast
cooling peltier system, its function can be activated
or deactivated.
The default setting for automatic deactivation of the
quick freeze is set to 30 minutes. The timer for
automatic deactivation can be set by an authorized
service technician.
UV disinfection
ON /OFF
If the instrument is equipped with the optional UV-
C disinfection system, its function can be activated
or deactivated.
The default setting for automatic deactivation is set
to 30 minutes. The timer can be set by an
authorized service technician.
Light
ON / OFF / AUTO
The light can be set on or off. If AUTO has been
selected, the light will switch off after the time
being set under Light duration.
Light duration
0 - 99 minutes