1. INTENDED USE
The floor standing open top cryostat TN60 is intended for rapid freezing and cutting
thin sections of frozen specimens for professional use in routine and research
laboratories in the fields of biology, medicine and industry.
The system of knife advance operates very reliable from 0.5 to 100 µm with trimming
function up to 750 µm. The quality of cutting of this microtome is increased by the
automatic retraction during the upstroke of the specimen, which avoids rubbing on the
disposable blades or microtome knives. This stops rapid deterioration of the disposable
blade or microtome knife.
2. SYMBOLS
Dangers, warnings and cautions are marked by this symbol
i
Special instructions regarding the operation of the
instrument are marked by this symbol
Mechanical components that can lead to injuries during
operation are marked with this symbol.
Symbol “flammable freezing sprays are prohibited”
3. SAFETY NOTES
The Tanner Scientific open top cryostat TN60 is provided with the following safety
features:
TN60
Hand wheel stop
•
Finger protection for Knife- and blade-holder
•
The institution which owns the unit and the persons working with the unit, servicing or
repairing it have the responsibility for a hazard-free use.