17 Manual_TN60_2021-09-001.00_EN_Tanner.docx
7.4 AUTOMATIC DEFROST, DISINFECTION AND POWER SAVING SETTINGS
Press
to open the Settings menu.
>
Chamber temp
-20 °C
Quickfreeze
Off
UV disinfection
Light
Auto
Select Program settings and press
to open the
menu for setting automatic defrost and power saving
times.
TIME TUE
13 05 05
Language
English
Display version
>
Program settings
5 µm
Enter PIN Code 5792.
PIN Code is usually restricted to service technicians or
instrument administrators
Enter PIN
0000
Select defrosting programs 1 to 3 and set a time for
the automatic defrosting programme.
(e.g. each day at 12:00 h and at 23:00 h in this
example)
>
Defr. Start 1
12:00
Defr. Start 2
23:00
Defr. Start 3
__:__
Disinf. Req.
__:__
Select temperature increase start and set a starting
time for the automatic power saving mode.
(e.g. start at 21:00 h in this example)
Disinf. delay
10 h
>
Incr. start
21:00
Incr. end
04:00
Incr. days
_ _ _ _ _ 5
6
Select temperature increase end and set a ending
time for the automatic power saving mode.
(e.g. end at 04:00 h in this example)
Disinf. delay
10 h
Incr. start
21:00
>
Incr. end
04:00
Incr. days
_ _ _ _ _ 5 6