8.12 ANTI-ROLL PLATE
Each instrument is equipped with
an anti-roll plate.
To adjust the anti-roll-plate
vertically, turn the wheel on the
base to the left (towards the blade)
or to the right (away from the
blade).
To remove the slices, swing the
anti-roll-plate to the left.
Verticale adjustment of anti-roll plate.
For an adjustment of the anti-roll-
plate (slit width), turn the screws on
the two sides of the plate. A
rotation of 30° corresponds to a
height adjustment of approx.
50µm.
Height adjustment of anti-roll plate.
The anti-roll-plate can be
exchanged with a few simple steps.
To do this, loosen the screws on the
anti-roll-plate-base and pull the
anti-roll-plate out of the guide. The
new anti-roll-plate can be used
simply by inserting it into the guide.
Before tightening the screws, please
check that the edge of the anti-roll-
plate is aligned with the blade!
Exchange of anti-roll-plate
i
For height adjustment of the anti-roll-plate it is recommended to
adjust the correct height during cutting.