Connecting to Your Computer

1.Slide the

impedance
meter down
from the top.

2.

Hang the

switch hanger
on the main
unit hook 

Specifications

Communication standard

Communication method

Signal speed

Data bit length

Parity

Stop bit

Flow control

Terminator

EIA RS-232C compatible

Asynchronous transaction

9600 bps

8 bits

None

1 bit

None

CR+LF

For Your Reference

A RS232C connector, D-sub
9 pin (female), is supplied
as standard with the main
unit.
Cables are not supplied and
need to be purchased
separately.

RS232C: straight D-sub 9 pin
(male) – D-sub 9 pin (female)
*Max. 3m

Data

Because data is output immediately following measurement regardless of the status of
the receiving side (personal computer, etc.) it is necessary to ensure that the receiving
side is always in receiving mode prior to conducting measurements.

(1) Output data format

Measurement data is output in the following format.
• Data segments are divided by commas (,).
• Terminators (the end of data) are CR (ASCll code 0DH), LF (ASCll code 0AH).

Model No.

MO

Gender

Trunk % fat

Estimated Waist Visceral fat level

Check Sum

XXXXXX

GE

1 or 2

FT

XX.X

WC

XXX.X

IF

XX.X

CS

XX

(2) Output Data Items

Item

Format

Content

Output Order

Standard Over 130cm

Control data

{0

Fixed at 16

1-5 byte variable length

1

1

Control data

˜0

Fixed at 1

1-5 byte variable length

2

2

Control data

˜1

Fixed at 1

1-5 byte variable length

3

3

Control data

˜2

Fixed at 1

1-5 byte variable length

4

4

Model number

MO

“AB-140”

8 byte fixed length

5

5

Gender

GE

1 or 2

1 byte fixed length (1: Male, 2: Female)

6

6

Trunk % fat

FT

XX.X

3-4 byte variable length, 1 decimal place (unit: %)

7

7

Waist circumference

WC

XXX.X

4-5 byte variable length, 1 decimal place (unit: cm)

8

X

Visceral fat level

IF

XX.X

3-4 byte variable length, 1 decimal place

9

8

Check sum

CS

XX

2 byte fixed length

10

9

Header

7

Storing the impedance meter

Always store the impedance meter in the main unit upon completion of
measuring (the impedance meter transfer mode will end and the
impedance meter will return to a state ready for use).

8

End measuring

Press the 

key on the main unit to turn off.

*If the main unit power is not turned off, it will automatically turn off after the

results have been displayed for approximately two minutes.

Check that the
Power LED is off

For Your Reference

Take care not to catch the body hair or skin of
the subject when moving the impedance meter
on the abdomen. Use both hands on either side
of the impedance meter, as shown in the
diagram.

Measuring

waist circumferences over 130cm (cont’d)

Summary of Contents for ViScan AB-140

Page 1: ... AB 140 Abdominal Fat Analyser Misuse of this device may result in a serious accident Please ensure that you read this Instruction Manual thoroughly prior to use and store it in a convenient location for easy reference N B 2008 TANITA Corporation All Rights Reserved AB1407601 0 ...

Page 2: ...pedance Analysis BIA method to directly measure the abdomen of a subject in the supine position With this method the previously difficult task of managing the health of people who are unable to stand or are otherwise disabled becomes possible The AB 140 also has the latest reactance technology functions which allows analyses to reflect cellular changes in the abdominal cavity and individual differ...

Page 3: ...transmission of disease when the meter is placed on the abdomen Use only the dedicated AC adapter and cord The use of other adapters cords may result in damage or fire Ensure disabled persons do not attempt to use the equipment by themselves A companion must perform the measurements Continually monitor both the subject and the equipment for anomalies Should an anomaly with the subject or the equip...

Page 4: ...allow to stand for at least 2 hours before using Wet the area of the abdomen to which the electrodes are to be attached and ensure correct positioning To wet lightly wipe the area with moistened cotton wool or a damp cloth Due to the influence of the state of the skin and body hair etc abnormally high readings and measurement errors may occur if the device is used without wetting the area Use in a...

Page 5: ...be accommodated by with the main unit shifts to calibration mode when pressed for 2 seconds or more Power key Press to turn on and off Measure key Battery level Press to begin measuring Charging LED Green light comes on when the AC adapter is connected light goes off when charging is completed Gender selection key Press to select subject gender Carry bag Distance measuring sensors Power LED Infrar...

Page 6: ...ies in accordance with the regulations of your local governing body 2 Loosen screws 3 Turn over to front side and remove the cover 1 Turn over Image only Please Charge the battery immediately after purchase The main unit cannot measure during charging Remove the AC adapter from the main unit when taking measurements moving the main unit charging is complete 4 Insert batteries 1 Insert the AC adapt...

Page 7: ... as shown in the diagram above As the distance sensor measures the breadth of the abdomen ensure that no clothes or elbows impinge on the 10cm area centred on the navel Do not allow the distance sensors to come into contact with the abdomen Refer to page 21 Waist Circumferences Over 130cm when contact is unavoidable due to a large waist or when the main unit cannot accommodate the size of the abdo...

Page 8: ...s been selected 5 Blue print in display illustrations indicates parts that illuminate or flash Blue print in display illustrations indicates parts that illuminate or flash 2 Once the gender has been selected the screen will change as shown At this time the red Laser On LED will light up and the laser in the main unit will operate To change the gender selection turn the power off then re select Mea...

Page 9: ...e laser directly towards the eyes as this may cause damage to eyesight Please Have the subject breathe normally while measurements are being taken Do not allow the distance sensors to come into contact with the abdomen Refer to page 21 Waist Circumferences Over 130cm when contact is unavoidable due to a large waist or when the main unit cannot accommodate the size of the abdomen For Your Reference...

Page 10: ...indicates parts that illuminate or flash Warning Do not direct the laser directly towards the eyes as this may cause damage to eyesight Please To wet wipe lightly with moistened cotton wool or other cloth Please Check that the green Power LED is illuminated when positioning the impedance meter The impedance meter Power LED will automatically turn off when more than approximately 10 minutes passes ...

Page 11: ...anger on the main unit hook Press the key on the main unit to turn on The initial screen is displayed This mode allows measurements to be easily taken when the waist of subject is too big to be accommodated by the main unit Place the main unit close to the supine subject as shown in the diagram below The main unit will switch off automatically after two minutes in the absence of any operation 1 Pr...

Page 12: ...ositioned close by remote controllers for televisions etc are used near the impedance meter or the main unit the batteries have expired the impedance meter is outside the infrared ray communication range see diagram at right the communication distance is changed due to the effect of lighting appliances 25cm Please To wet wipe lightly with moistened cotton wool or a damp cloth Please Check that the...

Page 13: ...Item Format Content Output Order Standard Over 130cm Control data 0 Fixed at 16 1 5 byte variable length 1 1 Control data 0 Fixed at 1 1 5 byte variable length 2 2 Control data 1 Fixed at 1 1 5 byte variable length 3 3 Control data 2 Fixed at 1 1 5 byte variable length 4 4 Model number MO AB 140 8 byte fixed length 5 5 Gender GE 1 or 2 1 byte fixed length 1 Male 2 Female 6 6 Trunk fat FT XX X 3 4 ...

Page 14: ...n be addressed by calibrating the unit thus allowing accurate measurements to be taken again To Calibrate Correctly Calibrate on a flat stable surface Ensure no objects are between the distance sensors Calibrate at a temperature of 15 25 C Press the key on the main unit to turn on The initial screen is displayed 1 Blue print in display illustrations indicates parts that illuminate or flash No obje...

Page 15: ...lacing Please contact a Sales Office for a new battery See the last page for contact details Access battery from the bottom Battery housing Caution Never switch off the power while is displayed indicating the unit is calibrating Calibration cont d The battery in the main unit is a consumable item The battery must be replaced when the charge soon runs out even after lengthy recharging Remaining bat...

Page 16: ...ing down Ensure the equipment is not exposed to any strong impact as this may cause damage Take care not to drop the equipment when removing from the carry bag as this may cause damage 30 How to Use 1 Remove the single screw from the bottom of the battery housing and push up the lid 2 Draw out the connector 3 Remove the battery 5 Connect the connector of the new battery to the main unit and insert...

Page 17: ...d through actively engaging in exercise and food restrictions Please see a medical practitioner for a medical diagnosis Very High 5 3 High No need for concern at present Continue with a balanced diet and an appropriate amount of exercise Average 1 18 0 20 0 13 0 15 0 1 0 6 0 2 6 5 12 5 4 15 5 17 5 6 20 5 and more Determination Interpretation Visceral fat levels measured by this unit It is possible...

Page 18: ... immediately Check the AC adapter connection Nothing is displayed when the power is turned on The Charging LED does not light up even when charging has commenced Display 34 As Needed Troubleshooting Problem Try This Please try the following prior to arranging repairs Measu ring Impedance meter error Buzzer does not sound or or is displayed Store the impedance meter in the main unit to reset the im...

Reviews: