18 

2.2 

Front Panel Control Functions 

 

2.2.1 

ITEM 

 

FUNCTION 

 
 

Supply Switch 

To isolate both supply leads from the set. 

 

ON/OFF 

 
 

Supply on Lamp 

To indicate supply is connected to the set  

 

 

 

 

and that supply fuse is healthy. 

 
 

 10 push-button 

Selects range for output milliammeter. 

 
 

HV OFF illuminated 

Indicates HV output off when lit.   Switches 

 

push-button (Green) 

off HV output when pressed. 

 
 

HV ON illuminated 

Indicates HV output on when lit.   Switches 

 

push-button (Red) 

on HV output when pressed. 

 
 

Input module  

Selects either 240/115V supply voltages 

 

 

 

 

houses mains fuse. 

 

Summary of Contents for PT15-10s

Page 1: ...DC Cable Test Set Type PT15 10s DESIGNED AND MANUFACTURED BY T R Test Equipment Limited 15 16 Woodbridge Meadows Guildford Surrey GU1 1BJ United Kingdom Telephone 01483 207428 e mail sales trtest com...

Page 2: ...2...

Page 3: ...he equipment that may be connected to a local earth Ensure that the product is properly grounded before making any connections to inputs or outputs Terminal ratings must be observed to prevent fire ha...

Page 4: ...4...

Page 5: ...5 CONTENTS 1 Description of Equipment 2 Operation of Equipment 3 Maintenance Instructions 4 Standard Accessories 5 Overall Performance Specification 6 Revision...

Page 6: ...6...

Page 7: ...1 Electrical Specification 1 2 Output Voltage Control 1 3 Overload Protection 1 4 High Voltage Output Cables 1 5 Metering 1 6 Automatic Earthing 1 7 Manual Earthing Using DP20 Discharge Earthing Prob...

Page 8: ...8...

Page 9: ...mA c The maximum currents available from the equipment are as the following table V supply kV DC mA DC R load C load 240V 15 0 500pF 2 14 7M 0 075 F 5 23 2 86M 0 15 F 8 4 1 78M 0 15 F 10 00 1 5M 0 22...

Page 10: ...F5A for 115V operation 1 3 2 The outputs from the instrument are protected by means of a fast acting magnetic overload relay which in turn controls an output contactor The overload is set for 12 15 m...

Page 11: ...milliamps dc The range change on the milliammeters is achieved by means of an non latching push button which is coloured white and located on the instrument s front panel The rest position is Range 1...

Page 12: ...ge Probe and More Importantly Possible Fatal Injury to the Operator 1 6 2 The Specification for DP20 Max discharge voltage kVdc 20kVdc Max discharge capacitance 12 F Max discharge energy 3 6kJ Max dis...

Page 13: ...the input and output sockets 1 9 Storage 1 9 1 The PT15 10S when not in use must be stored in a dry environment corrosion damage may occur to the HV components if stored in damp environment 1 9 2 Impo...

Page 14: ...14...

Page 15: ...15 2 OPERATION 2 1 Supply Voltage Selection 2 2 Front Panel Control Functions 2 3 Testing Procedures and Safety Precautions...

Page 16: ...16...

Page 17: ...til desired supply voltage is seen through the fuse drawer window c The PT15 10s is suitable for 2 different supply voltages as follows 240V or 115V d Replace fuse drawer once the desired voltage has...

Page 18: ...the set and that supply fuse is healthy 10 push button Selects range for output milliammeter HV OFF illuminated Indicates HV output off when lit Switches push button Green off HV output when pressed H...

Page 19: ...ype F 115V 5A Type F NEG HIGH VOLTAGE kV DC mA DC DANGER EARTH Negative kilovoltmeter Milliamp meter 10 push button Power ON OFF key switch Output ON push button Output OFF push button Fused mains inp...

Page 20: ...d to the cable sheath If the cable sheath is not earthed this may occur on cables which have just been installed a temporary earth spike 0 5M long may be used However this should be treated with cauti...

Page 21: ...ge regulator to the zero voltage position e Observe the slow decay in the output voltage which will indicate a healthy cable f After 30 seconds switch off the HV output The PT15 10s has an auto earth...

Page 22: ...ET TYPE PT15 10S POWER ON HV OUTPUT ON OFF INPUT 1ph 50 60Hz 240V 3A Type F 115V 5A Type F NEG HIGH VOLTAGE kV DC mA DC DANGER EARTH Cable Under Test Impedance Earth Connect to Low DP20 Earthing Disch...

Page 23: ...cking the brush condition as detailed above also check for signs of erosion on the regulator track Wipe clean with an alcohol moistened cloth If the track is badly eroded it should be cleaned with ver...

Page 24: ...om Its Housing 1 Remove the 4 fixing screws that hold the carrying strap to the unit s case 2 The unit can now be lifted from the case 3 To gain access to the internal components remove the aluminium...

Page 25: ...4 2 The following items are provided with the equipment a Mains input lead b 2 keys for mains ON OFF switch c 5 metre earth lead fitted with clip and tag d DP20 discharge earthing probe e 5 metre HT o...

Page 26: ...s than 10 Megohms between Mains Input to Frame 5 2 Applied Voltage Test Mains input to frame 2 kV RMS for 1 minute 5 3 Flashover test 5 flashovers at 15kV DC direct to earth 5 4 Normal voltage test 15...

Page 27: ...27...

Page 28: ...28 6 REVISION Product Type High Voltage DC Cable Test Set PT15 10s File PT15 10s manual v10 doc Author D Buckle Issue Date 10 28 02 17 Modified By K Mupotsa Checked By G Biggs Date 30 05 17...

Page 29: ...29 Drawings Required A2 000438 latest issue...

Reviews: