37
2.4.1
USB key
A standard USB flash memory key is used to store the results files. The unit has been tested
with keys up to 8Gb in size.
Insert the USB key into the upper USB socket on the unit before starting to test. Most USB
memory devices have an LED that indicates when the device is accessed. This will flash when
a result is written.
Do not unplug the USB key whilst data is being written – data loss will result.
2.4.2
Setting the clock
The output of the unit must be switched off to set the clock.
Set the STORE RESULT switch to SET CLOCK.
Use the
and
keys to select which digit to change.
Use the ADJ control to change the selected digit.
Use the red ON button to save the changes to the clock.
Use the green OFF button to cancel setting the clock.
Set the STORE RESULT switch to OFF.
2.4.3
Entering a comment
Each result is tagged with a comment when it is stored. The comment is entered by setting the
store result control to COMMENT and typing the comment using the optional USB keyboard.
The same comment is tagged to each result until the comment is updated or the comment is
erased. The comment is erased when the unit is switched off.
Any English language USB keyboard can be used with the CU-Ps.
The comment can be typed on the USB keyboard using the letters a-z, numbers 0-9 and
punctuation marks. A maximum of 59 characters may be entered for the comment.
The following keys are also used:
Shift + a-z
Enter capital letter.
Move cursor left/right.
Ins
Switch between overwrite and insert mode. By default, the unit will overwrite
the character to the right of the cursor when a new character is typed.
Switching to insert mode will shift characters of the right of the cursor right
when a new character is typed.
Enter
Accept the current comment.
hh:mm:ss dd-mm-yy