46
6.
PERFORMANCE SPECIFICATION
6.1.
Insulation Resistance at 1000V DC
The insulation resistance will not be less than 10 megohms between mains input to frame and
all isolated outputs, and all combinations of isolated output to isolated output.
6.2.
Applied Voltage Test
Mains input to frame: 2.0kV RMS for 1 minute.
All combinations of isolated output to isolated output and isolated output to frame: 1.0kV for 1
minute
6.3.
Accuracy of Instrumentation and associated circuit components
6.3.1
Main Output Ammeter
Range
Resolution
Accuracy
Trip current
20.00A
0.01A
0.6% rdg
5d
21.0A
50.00A
0.01A
0.6% rdg
5d
52.5A
100.0A
0.1A
0.6% rdg
5d
105A
200.0A
0.1A
0.6% rdg
5d
210.0A
6.3.2
Auxiliary Metering
Range
Resolution
Accuracy
Frequency (V)
40.00-99.99Hz
0.01Hz
0.2% rdg
2d
Frequency (I)
40.00-99.99Hz
0.01Hz
0.2% rdg
2d
Phase (V)
180.0°
0.1°
3 degrees
Phase (I)
180.0°
0.1°
3 degrees
AC Voltage
0-300.0Vac rms
0.1V
0.7% rdg
5d
AC Current
0-5.000Aac rms
0.001A
0.7% rdg
5d
6.3.3.
Timing System
Timer mode
Range
Resolution
Accuracy
Internal start
0-999.999s/
0-9999.99s/
0-99999.9s
1ms/10ms/100s
0.01% rdg
2d
Single contact
0.01% rdg
2d
Dual contact
0.01% rdg
2d
Current operated
0.01% rdg
4d
Pulse mode
500ms
1ms
2ms