If necessary the supply plug can be changed. Ensure that the plug is connected correctly. The
cores of the supply cable are marked to ensure correct phase rotation. The mains input lead
is marked as follows:
L1
Brown
L2
Black
L3
Grey
Neutral
Blue
Earth
Green and yellow
The unit requires a neutral connection (4 wire supply), and will not operate from a three wire
supply without the optional 200A-DSU delta-star converter.
It is essential that this unit is earthed.
Figure 1.1 Connections for mains input socket
1.2.3.1 Operation from a Single Phase Supply
If no three phase supply is available, it is possible to operate the 200A-3PH mk2 from a single
phase supply with reduced functionality. To connect the set to operate from a one phase
supply, connect L1, L2 and L3 to live, and N to neutral. When operating from a single phase
supply, all three outputs will be in phase.