tv2000-4000mkII-4000DC-man.doc Rev. 2.3 UK 9-15
This fine-tuning can be carried out at any time during
operation of the bath.
5.3
RS232C Control
A RS232C, bi-directional communication interface (300
to 38400 Baud) is integrated as standard, and offers the
possibility of computerised control.
5.4
Safety systems
A number of precautions is provided to ensure a safe
around the clock working of the bath. These features
protect the equipment, the bath fluid temperature and
the
workplace.
A
mechanical
over-temperature
protection thermostat (1) will automatically switch-off the
entire bath when its maximum value is exceeded. This
maximum temperature can be manually adjusted from
50°C up to 270°C.
A thermal protection of the stirrer mechanism (13) will
switch off the motor in case of malfunction. Both
thermostat and motor-fuse can easily be accessed. A
large number of integral electronic safety checks will
cause the bath to shut-down in case of electronic or
electrical error (11). Any kind of activated safety system
will be attended with an acoustic and visible alarm. On
the front display operating faults or component failures,
are reported as numbered ’errors’. In this way there is a
continual check of the proper functioning of the bath.