TROUBLE SHOOTING 

 

8.1 

Safety devices 

The  controller  is  equipped  with  a  number  of  safety 
devices  which  switch  off  the  bath  to  prevent  excessive 
temperature rise or malfunction. The safety devices are 
divided  in  mechanical  and  electronic  parts.  The 
mechanical  over-temperature  safety  thermostat  is 
located  near  the  mains  switch  and  consists  of  an 
adjustable  thermostat  and  a  red  arrow  shaped  lamp. 
The  desired  temperature  at  which  the  bath  should  be 
switched  off  can  be  set  by  turning  the  knob  with  a 
screwdriver or a coin.  
 

8.2 

Over-temperature safety thermostat  

When  the  thermostat  is  activated  the  entire  bath  is 
switched-off  (with  the  exception  of  the  controller)  the  
red  system  error  LED  will  light  up.  When  the 
temperature  of  the  bath  has  been  lowered  with 
approximately  20  degrees,  the  thermostat  re-sets  itself 
automatically, however to continue normal operation the 
bath has to be switched off and on again.  
 
The thermostat can be adjusted as follows:  
-Turn the thermostat fully clockwise. 
-Heat the bath to its proper temperature. Be aware that 
the safety thermostat is now only functioning at 270 °C. 
-Turn the thermostat gently counter clockwise, until  you 
hear  a  “click”.  Turn  the  knob  approximately  20  to  30° 
higher  (clockwise).  Switch  the  bath  Off  and  On  again. 
The bath is ready to operate safely. 
 

8.3 

Error messages 

There are two types of error messages: 

Fatal 

Non fatal 

In case of a fatal error the bath will be switched off. If a 
non  fatal  error  exists  the  error  will  be  repeated  after 
approximately 3 minutes.  
 

8.4 

Application errors  

Besides  the  electronic  errors  reported  on  the  display 
there are also practical errors.  
 

8.5 

Bath temperature does not become stable 

If  the  bath  temperature  does  not  stabilize  after  45 
minutes  after  set  point  has  been  reached  the  following 
points might cause this problem: 
·The operating temperature is around or below 30°C. To 
stabilize  the  bath  at  a  set-point  temperature  just  above 
the ambient temperature it is necessary to apply cooling 

Summary of Contents for PMT TV2000MKII

Page 1: ...21 19 41 E mail sales tamson com Website www tamson com VAT NL 80 66 34 984 B01 Bank account no Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code NL95 RABO 0160100046 Tamson Instruments...

Page 2: ...ITEMS 14 6 2 SAFETY THERMOSTAT 15 6 3 DISPLAY 15 6 4 QUICK START 16 7 MENU ITEMS 17 7 1 1 Menu item Set point 17 7 1 2 Menu item Offset 17 7 1 3 Menu item Max Power 17 7 1 4 Menu item Boost heater 17...

Page 3: ...VAT NL 80 66 34 984 B01 Bank account no Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code NL95 RABO 0160100046 Tamson Instruments bv tv2000 4000mkII 4000DC man doc Rev 2 3 UK 9 15 9 ADD...

Page 4: ...damage to the equipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect at the mains socket of the apparatus P...

Page 5: ...sonnel Avoid damage to the equipment or its accessories through incorrect operation 4 INSTALLATION 4 1 Important Tamson Instruments b v is not responsible for any consequential damage or harm caused b...

Page 6: ...d as long as the viscosity of the fluid is low enough at the operating temperature Following products can not be used as bath fluid Distilled water Aggressive fluids when in contact with stainless ste...

Page 7: ...oil 50 250 transparent 150cSt 50 C 122 F 6cSt 150 C 302 F F P 315 C 599 F Lifetime at 200 C 392 F is 1jr at 250 C 482 F 1200hrs when not polluted The oil used has a limited lifetime The type brand an...

Page 8: ...c Rev 2 3 UK 9 15 Using the bath for long periods at low temperatures will cause condensation on the glass window When operating at high temperatures the lid of the bath top plate and the window secti...

Page 9: ...UK 9 15 Use proper ventilation Heated bath fluids can cause toxic fumes Remove fumes from hot oil or bath medium Use fume cabinet or proper air ventilation When changing the bath fluid from water to o...

Page 10: ...to a waste container When the bath is working at high temperatures tubing and waste container must be chemical resistant and able to withstand high temperatures 230 C 446 F 4 3 2 Drain bath liquid Th...

Page 11: ...at lower temperatures it will be necessary to pass cooling liquid from a cooling circulator TLC10 or TLC15 or tap water through the cooling coil As an example to operate at 40 C 104 F with an ambient...

Page 12: ...temperature 5 1 Construction The TAMSON baths are constructed entirely from corrosion resistant materials such as stainless steel and PTFE Teflon A thick layer of rock wool between the inner bath and...

Page 13: ...nd the clock working of the bath These features protect the equipment the bath fluid temperature and the workplace A mechanical over temperature protection thermostat 1 will automatically switch off t...

Page 14: ...6 1 Overview menu items Set point Offset press 5 00 5 00 C resolution 0 01 C Max Power press low med hi max Boost heater press on off Time const press fast medium slow precise PID parameter PID set 1...

Page 15: ...d on again When switching off wait 5 seconds before switching on again 6 3 Display 1 Temperature readout 2 Applied percentage of power 3 Operating mode 4 Indicator alarm high alarm low control stable...

Page 16: ...start operating the bath in a quick way do the following Fill the bath with liquid as indicated place the power plug connect to mains socket switch on the bath using the mains switch select appropria...

Page 17: ...th temperature It can be set in steps of 0 01 C 7 1 2 Menu item Offset The temperature displayed can be increased or decreased with an offset ranging from 5 00 down to 5 00 in steps of 0 01 C The offs...

Page 18: ...gh alarm Optional min SP to max SP resolution 0 1 C No hardware connected display function only Optional 7 1 9 Menu item PID parameter PID set 1 First set of parameters PID set 2 Second set of paramet...

Page 19: ...et of 0 005 C The value of 0 005 C is added to the selected SP value The SP can be selected with 0 01 C accuracy 7 1 14 Menu item Restart Restarts system and activates tuning 7 2 PID settings Default...

Page 20: ...NL 80 66 34 984 B01 Bank account no Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code NL95 RABO 0160100046 Tamson Instruments bv tv2000 4000mkII 4000DC man doc Rev 2 3 UK 9 15 7 3 TV400...

Page 21: ...of the Ziegler Nichols method described below The PID parameters can also be determined with the use of the Ziegler Nichols method described below With the process at its normal process value Set the...

Page 22: ...the temperature of the bath has been lowered with approximately 20 degrees the thermostat re sets itself automatically however to continue normal operation the bath has to be switched off and on again...

Page 23: ...irculating system is incapable to mix the bath fluid thoroughly which may result in poor stability Power is limited to a very low level Position of PT100 has been changed The tip of the PT100 must be...

Page 24: ...hen PV is in the region of 100 C PV 0 C StableLimitHigh is defined as 0 05 When the average PV deviation calculated from a sliding window of 255 seconds is smaller than StableLimitHigh the ST command...

Page 25: ...decimal value with 7 decimal digits e g 12 345342C works for degrees Centigrade and degrees Fahrenheid Only first three are significant When reading is in degree Fahrenheit character F indicates the...

Page 26: ...lete TV2000MKII 25T0210 25T0230 Heater 1500W 25T0210 25T0210 Boost Heater 25T1344 Capacitor 5 5uF 24T3312 Capacitor 10uF 24T8081 Motorfuse 0 3 Amp 24T8080 Motorfuse 0 6 Amp 24T8541 Mains switch 25T124...

Page 27: ...an doc Rev 2 3 UK 9 15 9 ADDENDUM TECHNICAL REFERENCE 9 1 Technical Drawing TV2000 TV4000MKII Motor fuse Heater Boost Over temperature Thermostat Fan PT100 Neutral RS232 female Flatcable powerboard He...

Page 28: ...c Rev 2 3 UK 9 15 9 2 Technical drawing TV4000MKII DC Motor fuse Heater Boost Over temperature Thermostat Fan PT100 Neutral RS232 female Flatcable powerboard Level detect 12V DC PE Phase PE Mains Indi...

Page 29: ...e 2006 95 EC relative to low voltage The products are in conformity with the following specifications EMC 2004 108 EG Conducted emission EN55016 2 1 EN61326 A1 Radiated emission EN55016 2 3 EN61326 A1...

Page 30: ...every kind and nature including warranties of merchantability and fitness for a particular purpose in respect to the use or suitability of the product Nothing is intended as a recommendation for uses...

Reviews: