USER MANUAL PART NUMBER 50-02-07-212/5

This medical device is compliant with:

IEC 60601.1 3rd edition Medical electrical equipment safety and essential performance

IEC 60601.1.11 Home healthcare environment

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic Environment - Guidance

Electrostatic 

discharge (ESD) 

IEC 61000-4-2

± 6kV contact

± 8kV air

± 6kV contact

± 8kV air

Floors should be wood, concrete or ceramic tile. If floors 

are covered with synthetic material, the relative humidity 

should be at least 30%.

Electrical fast 

transient/burst IEC 

61000-4-4

± 2 kV For mains supply lines

± 1kV For input/output lines

± 2 kV For mains supply lines

± 1kV For input/output lines

Mains supply quality for the mains adapter should be 

that of a typical commercial and/or hospital environment.

Surge

IEC61000-4-5

± 1kV line(s) to line

± 1kV line(s) to line

Mains supply quality for the mains adapter should be 

that of a typical commercial and/or hospital environment.

Voltage dips, short 

interruptions and 

voltage variations 

on mains supply 

IEC 61000-4-11

<5%Ur (>95%Ur) for 0.5 cycle 

<5%Ur (>95%Ur) for 0.5 cycle 

Mains supply quality for the mains adapter should be 

that of a typical commercial and/or hospital environment. 

In the event of a mains interruption the ATTIVO™ 

system will automatically use internal battery power, 

unless the battery is exhausted.

40%Ur (60% dip in Ur) for 5 cycles

40%Ur (60% dip in Ur) for 5 cycles

70%Ur (30% dip in Ur) for 25 cycles 70%Ur (30% dip in Ur) for 25 cycles
>5%Ur (>95% dip in Ur) for 5 secs

>5%Ur (>95% dip in Ur) for 5 secs

Mains frequency 

(50/60Hz) magnetic 

field IEC61000-4-8

3 A/m

3 A/m

Mains frequency magnetic fields should be at levels 

characteristic of a typical location in a typical commercial 

and/or hospital environment.

Note: Ur is the A.C. mains voltage prior to application of the test level.

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)

Immunity 

Test

IEC 60601 

Test Level

Compliance 

Level

Electromagnetic Environment - Guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V rms

150 kHz ~ 80 MHz

3 V/m

80 MHz ~

2.5 GHz

3 V rms

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the 

ATTIVO™ including cables, than the recommended separation distance calculated from the equation 

appropriate to the frequency of the transmitter.

Recommend separation distance

d = 1.2 √P

d = 1.2 √P 80 MHz to 800 MHz

d = 2.3 √P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the 

transmitter manufacturer and d is the recommended separation

distance in meters (m).

Field strengths from fixed RF transmitters as determined by an electromagnetic site

survey,

a

 should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and 

people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, 

and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromag

-

netic site survey should be considered. If the measured field strength in the location in which the 

ATTIVO™ is used exceeds the applicable RF compliance 

level above, the ATTIVO™ should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, 

such as reorienting or relocating the ATTIVO™.

b

 Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.

Recommended Separation Distance Between Portable and Mobile RF Communications Equipment and the ATTIVO™

The ATTIVO™ is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the ATTIVO™ can help 

prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and 

the ATTIVO™ as recommended below, according to the maximum output power of the communications equipment.

Output Power of Transmitter in

Watts (W)

Separation distance according to frequency of transmitter in Meters (m)

150 kHz to 80 MHz d = 1.2 √P

80 MHz to 800 MHz d = 1.2 √P

800 MHz to 2.5GHz d = 2.3 √P

0.01

0.01

0.01

0.01

0.12

0.12

0.12

0.12

0.12

0.12

0.12

0.12

0.23

0.23

0.23

0.23

0.1

0.1

0.1

0.1

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in Meters (m) can be estimated using the 

equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to the 

transmitter manufacturer. Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies. Note: These guidelines may not 

apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

02/2017

Talley Group Limited

Premier Way, Abbey Park Industrial Estate, Romsey, Hampshire, SO51 9DQ England

TEL

+

44(0)1794 503500   

FAX

:  

+44(0)1794 503555

E-MAIL

:  

[email protected]

www.talleygroup.com

Summary of Contents for ATTIVO

Page 1: ...ATTIVO Dynamic Seating System User Guidelines reducing avoidable harms ...

Page 2: ...This is a statement that alerts the user to the possibility of serious injury or other adverse reactions with the use or misuse of the device This is a statement that alerts the user to the possibility of a problem with the system associated with its use or misuse IP Ingress Protection Pump only 2 Protection against fingers or other object not greater than 80mm in length and 12mm in diameter 2 Pro...

Page 3: ...Power unit carry case l TS209 B A S E SEQUENTIAL cushion or B A S E Recliner Mat The following accessories spares are also available l 12V Vehicle adaptor charger l Spare mains lead with integral power supply 15V Mains adapter FW7362M 15 Cautions and Warnings CONTRAINDICATIONS FOR USE Alternating pressure therapy should not be used for patients with unstable fractures gross oedema burns or an into...

Page 4: ...ct fuse used l The cushion power unit and power adapter should be cleaned between patient use please refer to Care and Maintenance section for all warnings and cautions l The power unit battery is installed inside the unit and is not accessible by users Battery replacement should only be carried out by qualified Talley service personnel l The power unit must only be used with the approved power ad...

Page 5: ...e the battery is fully charged when needed A fully discharged battery will take a number of hours to fully charge 6 Standby mode is automatically invoked when the power is switched on Press the RUN STOP button to initialise and run the seating system When the power unit has finished initialising the pump screen will display ATTIVO together with the battery status and the comfort setting Fig 2 7 Th...

Page 6: ...isplay screen The MUTE button can also be used to force power off in stand by mode DATA The information menu can be activated by pressing the DATA button for 5 or more seconds while pressing the RUN STOP button then releasing the RUN STOP button while continuing to press the DATA button for a few seconds Press the DATA button when ready to exit the information menu NB Used for accessing informatio...

Page 7: ...acement should only be carried out by qualified service personnel Contact Talley or authorised dealer for battery replacement service Care and Maintenance COVER Always keep the cushion cover as clean as is practicable The material is waterproof and vapour permeable l Inspect top cover for signs of damage or wear which could result in the contamination of the interior e g tears holes damage to seam...

Page 8: ...adapter in accordance with the local regulations including WEEE requirements The power unit power adapter should be cleaned between patient use as a minimum In line with the MHRA Medical Device Alert MDA 2013 019 Talley advises customers to use pH neutral high level disinfectant cleaning products to sanitise reusable medical devices to prevent damage to materials and the degradation of plastic sur...

Page 9: ...ntee The intended design life is 5 years if fully serviced Fault Finding All sounders can be silenced and messages cleared by pressing the MUTE button once Should any fault occur press the MUTE button to reset the power unit If fault remains re occurs contact Talley WARNING INDICATORS NO BATTERY indicates battery is not correctly fitted The power unit will only operate using the power adapter whil...

Page 10: ...mensions 430mm x 430mm x 70mm 1080mm x 430mm x 70mm Weight 1 9 kg 3 9 kg The above mains adapter is considered part of the ME equipment 12V vehicle power adapter is available as an optional accessory 99 01 14 05 The ATTIVO power unit must only be used with the specific external power adapters as supplied by Talley EMI fault indicates that the unit detects the pressure sensor amplifier is adversely...

Page 11: ...nterfering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipment to an outlet on a circuit different from that to which the interfering equipment was connected Information regarding Electro Magnetic Compatibility EMC according to IEC60601 1 2 With the increa...

Page 12: ...urer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not...