It is the customer’s responsibility to ensure the following prior to collection:
l
the system is cleaned of any obvious contaminants.
l
contamination status is documented.
l
assistance is given to Talley personnel to bag the equipment if the mattress has been in
a known or suspected infectious environment.
TRANSPORT AND STORAGE
Handle with care. Please report instances of damage or impact to Talley Service Department.
Transport
–25 °C without relative humidity control; and
+70 °C at a relative humidity up to 93 %, non-condensing.
An atmospheric pressure range of 700 hPa to 1 060 hPa.
Suitable for all standard modes of transport when in the correct packaging.
OPERATIONAL CONDITIONS
A temperature range of +5 °C to +40 °C;
A relative humidity range of 15% to 93%, non-condensing; and
Operational Atmospheric Pressure: 700 hPa to 1060 hPa
Suitable for pollution degree 2
Operational altitude ≤ 2 000 m
IP Rating: IP22 pump only
MANUFACTURER’S GUARANTEE
The power unit and cushion are covered by a 24 month manufacturer’s guarantee. The
intended design life is 5 years if fully serviced.
Fault Finding
All sounders can be silenced and messages cleared by pressing the MUTE button once.
Should any fault occur, press the MUTE button to reset the power unit. If fault remains/
re-occurs, contact Talley.
WARNING INDICATORS
NO BATTERY
– indicates battery is not correctly fitted. The power unit will only operate using
the power adapter whilst this message is displayed. Contact Talley or authorised dealer to
check battery and installation.
LOW BATTERY
(only appears during battery operation) – the power unit will continue to
run whilst this warning is displayed. Press the MUTE button to silence the sounder and clear
the message. Note that the system will automatically shut down when the battery is fully
discharged. Plug into a power source to charge.
LOW PRESSURE
– indicates pressure has fallen below minimum allowable levels. Power
unit will continue to run whilst these messages is displayed. Check that the hose is connected
to the power unit correctly. Check that the internal tubing and bellows are connected and
that there are no punctures. Press the MUTE button to clear the message and to silence the
sounder. If the fault re-occurs, contact Talley.