background image

8

It is the customer’s responsibility to ensure the following prior to collection:

the system is cleaned of any obvious contaminants.

contamination status is documented.

assistance is given to Talley personnel to bag the equipment if the mattress has been in 

a known or suspected infectious environment.

TRANSPORT AND STORAGE

Handle with care. Please report instances of damage or impact to Talley Service Department.

Transport

–25 °C without relative humidity control; and

+70 °C at a relative humidity up to 93 %, non-condensing.

An atmospheric pressure range of 700 hPa to 1 060 hPa.

Suitable for all standard modes of transport when in the correct packaging.

OPERATIONAL CONDITIONS

A temperature range of +5 °C to +40 °C;

A relative humidity range of 15% to 93%, non-condensing; and

Operational Atmospheric Pressure: 700 hPa to 1060 hPa

Suitable for pollution degree 2

Operational altitude ≤ 2 000 m

IP Rating: IP22 pump only

MANUFACTURER’S GUARANTEE

The power unit and cushion are covered by a 24 month manufacturer’s guarantee. The 

intended design life is 5 years if fully serviced.

Fault Finding

All sounders can be silenced and messages cleared by pressing the MUTE button once.  

Should any fault occur, press the MUTE button to reset the power unit.  If fault remains/

re-occurs, contact Talley.

WARNING INDICATORS

NO BATTERY

 – indicates battery is not correctly fitted. The power unit will only operate using 

the power adapter whilst this message is displayed. Contact Talley or authorised dealer to 

check battery and installation.

LOW BATTERY

 (only appears during battery operation) – the power unit will continue to 

run whilst this warning is displayed. Press the MUTE button to silence the sounder and clear 

the message. Note that the system will automatically shut down when the battery is fully 

discharged. Plug into a power source to charge.

LOW PRESSURE 

– indicates pressure has fallen below minimum allowable levels. Power 

unit will continue to run whilst these messages is displayed. Check that the hose is connected 

to the power unit correctly. Check that the internal tubing and bellows are connected and 

that there are no punctures.  Press the MUTE button to clear the message and to silence the 

sounder. If the fault re-occurs, contact Talley.

Summary of Contents for ATTIVO

Page 1: ...ATTIVO Dynamic Seating System User Guidelines reducing avoidable harms ...

Page 2: ...This is a statement that alerts the user to the possibility of serious injury or other adverse reactions with the use or misuse of the device This is a statement that alerts the user to the possibility of a problem with the system associated with its use or misuse IP Ingress Protection Pump only 2 Protection against fingers or other object not greater than 80mm in length and 12mm in diameter 2 Pro...

Page 3: ...Power unit carry case l TS209 B A S E SEQUENTIAL cushion or B A S E Recliner Mat The following accessories spares are also available l 12V Vehicle adaptor charger l Spare mains lead with integral power supply 15V Mains adapter FW7362M 15 Cautions and Warnings CONTRAINDICATIONS FOR USE Alternating pressure therapy should not be used for patients with unstable fractures gross oedema burns or an into...

Page 4: ...ct fuse used l The cushion power unit and power adapter should be cleaned between patient use please refer to Care and Maintenance section for all warnings and cautions l The power unit battery is installed inside the unit and is not accessible by users Battery replacement should only be carried out by qualified Talley service personnel l The power unit must only be used with the approved power ad...

Page 5: ...e the battery is fully charged when needed A fully discharged battery will take a number of hours to fully charge 6 Standby mode is automatically invoked when the power is switched on Press the RUN STOP button to initialise and run the seating system When the power unit has finished initialising the pump screen will display ATTIVO together with the battery status and the comfort setting Fig 2 7 Th...

Page 6: ...isplay screen The MUTE button can also be used to force power off in stand by mode DATA The information menu can be activated by pressing the DATA button for 5 or more seconds while pressing the RUN STOP button then releasing the RUN STOP button while continuing to press the DATA button for a few seconds Press the DATA button when ready to exit the information menu NB Used for accessing informatio...

Page 7: ...acement should only be carried out by qualified service personnel Contact Talley or authorised dealer for battery replacement service Care and Maintenance COVER Always keep the cushion cover as clean as is practicable The material is waterproof and vapour permeable l Inspect top cover for signs of damage or wear which could result in the contamination of the interior e g tears holes damage to seam...

Page 8: ...adapter in accordance with the local regulations including WEEE requirements The power unit power adapter should be cleaned between patient use as a minimum In line with the MHRA Medical Device Alert MDA 2013 019 Talley advises customers to use pH neutral high level disinfectant cleaning products to sanitise reusable medical devices to prevent damage to materials and the degradation of plastic sur...

Page 9: ...ntee The intended design life is 5 years if fully serviced Fault Finding All sounders can be silenced and messages cleared by pressing the MUTE button once Should any fault occur press the MUTE button to reset the power unit If fault remains re occurs contact Talley WARNING INDICATORS NO BATTERY indicates battery is not correctly fitted The power unit will only operate using the power adapter whil...

Page 10: ...mensions 430mm x 430mm x 70mm 1080mm x 430mm x 70mm Weight 1 9 kg 3 9 kg The above mains adapter is considered part of the ME equipment 12V vehicle power adapter is available as an optional accessory 99 01 14 05 The ATTIVO power unit must only be used with the specific external power adapters as supplied by Talley EMI fault indicates that the unit detects the pressure sensor amplifier is adversely...

Page 11: ...nterfering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipment to an outlet on a circuit different from that to which the interfering equipment was connected Information regarding Electro Magnetic Compatibility EMC according to IEC60601 1 2 With the increa...

Page 12: ...urer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not...

Reviews: