During operation, any rattling or ticking sounds may indicate a loose screw on the
tray, a tray attachment or an accessory. All accessories should be sufficiently tightened
in place before starting the unit.
Error Code
Software test
Cause
E04
unit overloaded
maximum load exceeded
Press the standby button to clear this error. Be sure the load is within the maximum
load capacity before restarting the unit. If the E04 error persists, switch the unit off and
contact your Talboys representative for repairs.
Error Code
Software test
Cause
E03
drive system failure ceased motor
drive belt broken
mechanical obstruction
In the event of someone/something inadvertently touching the tray while the unit
is running, the unit will display an E03 error message, and automatically go to the
horizontal “home” position and restart.
In the event of a mechanical obstruction, the unit will attempt to auto “home” and
restart four (4) times and then stop running, displaying an E03 error message. Press
the standby button to clear this error and remove the mechanical obstruction.
If the E03 error persists the reason may be a ceased motor or broken drive belt
and should not be addressed by the end user. Switch the unit off and contact your
Talboys representative for repairs.
Error Code
Cause
E02
RTD shorted or temperature below 0°C (32°F)
E01
RTD open or temperature over 100°C (212°F)
The E02 and E01 errors should not be addressed by the end user. Switch the unit
off and contact your Talboys representative for repairs.