background image

 

Model 1700 / 1750 Rotary Platform Abraser 

19 

Operating Instructions ver. 1.0 

 

 

p/n 135505 

 

 

10.  After  completion  of  test  cycles,  remove  and 

clean test specimen. 

11.  Evaluate  and  record  intermediate  or  final 

values based on method selected for reporting 
test result. For weight loss, weigh specimen to 
nearest  0.1  mg.  For  visual  comparisons, 
examine test surface and note characteristics. 

 

 

WARNING:

 

When  testing  flexible  materials,  trim 

any  excess  material  overhanging  the  specimen 
holder  or  hold-down  ring.  Otherwise,  the  material 
may  rub  on  the  abraser  housing  and  cause 
damage. 

 

NOTE:

 

During  usage,  the  abrasive  wheels  will  break 

down  causing  the  diameter  to  change.  Wheels  must  be 
replaced  when  the  diameter  wears  down  to  the  wheel 
label (44.4 mm).

 

 

NOTE:

 

The  PAUSE  INTERVAL  will  stop  the  test  at 

predetermined number of cycles and is useful for tests that 
require  the  abradant  to  be  refreshed  or  to  evaluate  the 
specimen during the test (e.g.  visual inspection of wear-
through 

to 

the 

substrate 

or 

periodic 

weight 

measurements). 

 

 

CALCULATION OF RESULTS 

 

The  following  are  commonly  used  methods  for 
reporting abrasion test results. For comparable and 
reproducible  tests,  it  is  recommended  that  all 
testing  be  performed  under  conditions  covered  by 
an established test procedure. 

 

Weight Loss 

The  weight  loss  test  method  is  a  quantifiable 
method  that  records  the  weight  loss  of  the  test 
specimen  due  to  abrasion.  This  method  is 
recommended when the results are to be compared 
with  those  of  similar  materials  having  nearly  the 
same specific gravity. 

Weigh the test specimen before and after testing to 
obtain  the  initial  and  final  weight  values.  The 
difference between these two values, 

F

Total

, will be 

the weight loss. 

𝐹

𝑡𝑜𝑡𝑎𝑙

= 𝐴 − 𝐵

 

Where,  A = weight of test specimen before abrasion, mg 

 

B = weight of test specimen after abrasion, mg

 

Use  the  following  formula  to  calculate  average 
mass loss, 

F

m

, in milligrams (mg) per 100 cycles: 

𝐹

𝑚

=

𝐹

𝑡𝑜𝑡𝑎𝑙

𝑛

𝑥100

 

Where,  n = total number of cycles

 

When  performing  the  weight  loss  method,  loose 
particulate may adhere to specimens during testing. 
It is critical that test specimens are cleaned prior to 
weighing. If a sample has been wet tested or if an 
indicator  has  been  used,  ensure  the  specimen  is 
thoroughly  dried.  If  static  electricity  affects  the 
specimen,  a  static  eliminator  may  be  sprayed  on 
both sides prior to taking the final measurement. 

Taber  Wear  Index  may  also  be  used  to  present 
weight loss data. This measurement represents the 
loss in weight in milligrams per thousand cycles of 
abrasion

 

for a test performed under a specific set of 

conditions. The lower the wear index, the better the 
abrasion resistance quality of the material. 

𝑇𝑎𝑏𝑒𝑟𝑊𝑒𝑎𝑟 𝐼𝑛𝑑𝑒𝑥 =  

𝐹

𝑡𝑜𝑡𝑎𝑙

 𝑥 1000 𝑐𝑦𝑐𝑙𝑒𝑠

𝑛

 

For example: if a specimen is abraded 5000 cycles 
and has a weight loss of 500 mg, the wear index will 
be  100.  Likewise,  a  material  that  withstood  500 
cycles  of  abrasion  and  had  a  weight  loss  of  100 

Summary of Contents for 1700

Page 1: ...Taber Rotary Platform Abrasion Tester Model 1700 1750 Operating Instructions...

Page 2: ......

Page 3: ...ntanglement hazard Do not place body parts or objects in the area surrounding the specimen holder and wheels as this may cause injury and or damage the equipment WARNING To minimize airborne particula...

Page 4: ...CIRCUMSTANCES SHALL TABER BE LIABLE FOR ANY INJURIES DAMAGE TO OR REPLACEMENT OF OTHER PRODUCTS OR PROPERTY COSTS FOR RECOVERING REPROGRAMMING OR REPRODUC ING ANY PROGRAM OR DATA USED WITH THE PRODUCT...

Page 5: ...t the wheels traverse a complete circle on the specimen surface revealing abrasion resistance at all angles relative to the weave or grain of the material It is important to recognize there may be sev...

Page 6: ...t the vacuum unit to the Abraser 6 Plug the female connector of the Abraser power cord into the power inlet located on the rear of the instrument Connect the instrument to a 115V 60 Hz or 230V 50 Hz c...

Page 7: ...button and then select the desired option After making your selection press OK NOTE Options shown below in bold text are factory default settings Language English Mandarin Japanese Spanish French Ger...

Page 8: ...test parameters When the instrument is not operating press the appropriate button and use the dial or odometer to set the value After making your selection press OK Test Cycles The Abraser can be pre...

Page 9: ...shown for the correct specimen holder To change between the left or right press the arrow button on the side of the display screen or press the toggle switch at the top status bar LEFT or RIGHT will...

Page 10: ...uum suction for accurate standardization and control of test procedures WARNING Do not use the vacuum for wet tests To minimize risk of shock disconnect the electrical cord to the vacuum unit NOTE Whe...

Page 11: ...be refreshed at intervals during the test NOTE Do not exceed 50 resurfacing cycles on an S 11 refacing disc If more than 50 cycles are required to refresh the wheels use additional refacing discs Vac...

Page 12: ...Operating Instructions ver 1 0 p n 135505 Display Option The Rotary Platform Abrasion tester can be programmed to show Cycles or Time Both Cycles and Time can be displayed as ascending completed Cycle...

Page 13: ...e way onto the mounting hub Release the push button The hub retaining lip will secure the wheel until it is disengaged Genuine Taber abrading wheels are supplied in pairs A close inspection of the whe...

Page 14: ...al counterweight kit is available when less than 250 g load is required Vacuum System Models 1700 1750 include a Taber vacuum unit that includes a heavy duty motor and life lubricated sealed bearings...

Page 15: ...larly defined intervals Do not confuse coloring of the wheels with clogging as this may simply be due to the wheel faces becoming coated with fine abrading dust a condition that must be expected It is...

Page 16: ...and mount the test specimen Select CONTINUE to continue testing or REFACE to perform a second refacing NOTE The S 11 refacing disc is used only once for a maximum of 50 cycles Never reface wheels wit...

Page 17: ...ter edge of each wheel Manually turn the nut to rotate the motor shaft Reverse the wheels on the arbor and round the opposite corners Only the extreme sharpness of the edges should be removed A radius...

Page 18: ...plate and nut and finally tighten the hold down ring Avoid tightening to excess so as not to wrinkle the specimen Trim off any excess material around the edges WARNING To prevent damage to the instru...

Page 19: ...wear life under changing atmospheric conditions as in automobile and aircraft materials the atmosphere of the laboratory may be controlled accordingly NOTE If an environmentally controlled test labor...

Page 20: ...y 3 Secure specimen to Abraser specimen holder 4 Select and mount the auxiliary weights if necessary 5 Adjust the vacuum nozzle gap 6 Using the Abraser touchscreen enter the appropriate test parameter...

Page 21: ...iable method that records the weight loss of the test specimen due to abrasion This method is recommended when the results are to be compared with those of similar materials having nearly the same spe...

Page 22: ...cific gravities a correction factor must be applied to give a true indication of the wear resistance In doing this the aluminum material evidences a much greater volume loss as seen by comparison of t...

Page 23: ...es or more it is often attributed to procedural errors or an instrument that is out of calibration or one of the factors mentioned below Physical properties of the material such as hardness or resilie...

Page 24: ...for unpaired data at a probability level chosen prior to the testing series If bias is found either its cause must be found and corrected or future test results must be adjusted in consideration of t...

Page 25: ...5505 HELP At the bottom of the display screen is the HELP button Pressing this will provide details on the instrument USB Port Incorporated into the Model 1700 1750 Abraser is a USB port This is used...

Page 26: ...he holder and lift straight upward To install the specimen holder align the set screws with the flats on the motor shaft Cleaning Vacuum Pick Up Nozzle Abraded material and debris may collect in the v...

Page 27: ...14120 USA Phone 716 694 4000 Fax 716 694 1450 E mail sales taberindustries com WebSite www TaberIndustries com www OrderTaber com Part Description 133338 Quick Release Wheel Hub Collet Kit replacement...

Reviews: