background image

 

Model 1700 / 1750 Rotary Platform Abraser 

13 

Operating Instructions ver. 1.0 

 

 

p/n 135505 

 

NOTE:

 A gap of 3 mm has been found to be effective on 

most materials. 

 

NOTE:

  To facilitate  setting  the  nozzle  gap,  a  shim may 

be  inserted  between  the  vacuum  nozzle  and  specimen. 
Take care not to damage the specimen surface. 

Vacuum Level

 

– The vacuum suction level should 

be  set  high  enough  to  remove  debris  from  the 
specimen  surface,  but  not  lift  flexible  samples.  A 
visual  inspection  of  the  wear  path  can  usually 
indicate  if  debris  is  adhering  to  the  specimen 
surface  and  the  vacuum  level  (or  vacuum  nozzle 
height)  needs to be  adjusted. For test control  and 
standardization  of  vacuum  suction,  the  vacuum 
level is calibrated to values of 50 

– 100%. 

 

NOTE:

  The  amount  of  vacuum suction  is  influenced  by 

the type of material being tested and height of the nozzle 
above the specimen surface. 

Vacuum Nozzle Inserts

 

– The Model 1700 / 1750 

includes  replaceable  vacuum  nozzle  inserts  (front 
nozzle is short, rear nozzle is long). The nominal 8 
mm  are  standard  for  most  tests.  The  optional  11 
mm  nozzle  inserts  are  typically  required  for 
materials that are evaluated using a change in haze 
(e.g. ASTM D1044). 

       

 

 

 

REFACING WHEELS 

 

Taber  abrasive  wheels  possess  properties  that 
cause  the  working  faces  to  disintegrate  slowly  as 
they are used, continually exposing fresh abrasive 
grains  to  the  specimen  being  tested.  However, 
depending on the material being evaluated, debris 
from the test specimen may  clog or  adhere to  the 
wheel  surface.  If  this  occurs,  the  wheels  must  be 
cleaned (refaced) at regularly defined intervals. Do 
not confuse coloring of the wheels with “clogging” 
as  this  may  simply  be  due  to  the  wheel  faces 
becoming  coated  with  fine  abrading  dust,  a 
condition that must be expected. 

It is good practice to standardize the abradant prior 
to each test. Failure  to reface the  wheel surfaces, 
may introduce variation into the test results. 

 

NOTE:

 

Wear  life  of  Taber  abrasive  wheels  varies 

depending on test load, and the surface texture / frictional 
characteristics  of  the  material  being  tested.  Under 
ordinary conditions, a wheel set should last from 10,000 
to  50,000  wear  cycles.  Abnormally  short  wheel  life  may 
indicate the use of the wrong wheel, load, or procedure for 
the material tested, or too frequent refacing. 

 

NOTE:

 

Wheels can be used until they are worn down to 

their minimum  diameter  of  44.4  mm  as  indicated  by  the 
wheel label. 

 

Refacing Resilient Wheels (Calibrase) 

The  S-11  Refacing  Disc  are  used  to  resurface  all 
Calibrase  wheels  with  the  exception  of  CS-10F 
when they are used for evaluating change in haze 
for  transparent  materials.  In  this  case,  CS-10F 
wheels  are  refaced  using  the  ST-11  Refacing 
Stone. 

1.  Mount the wheels to the abraser. 

2.  Mount  an  S-11  Refacing  Disc  (or  ST-11 

Refacing  Stone)  on  the  SH-125  specimen 
holder. Use the clamp plate and nut, and hold-
down ring to secure the S-11. The clamp plate 
and clamp ring  are not required for the  ST-11 
refacing stone. 

3.  Mount the auxiliary weights for the load that will 

be used for testing. 

4.  Lower  the  abrading  wheels  onto  the  refacing 

medium. 

5.  Adjust  the  vacuum  pick-up  nozzle  height  to 

within 3 

– 6 mm of the refacing disc. See note 

on page 14 for CS-10F wheels. 

6.  Program  the  appropriate  number  of  cycles 

according  to  Table  1  (or  as  specified)  and 
subject the wheels to refacing. 

Summary of Contents for 1700

Page 1: ...Taber Rotary Platform Abrasion Tester Model 1700 1750 Operating Instructions...

Page 2: ......

Page 3: ...ntanglement hazard Do not place body parts or objects in the area surrounding the specimen holder and wheels as this may cause injury and or damage the equipment WARNING To minimize airborne particula...

Page 4: ...CIRCUMSTANCES SHALL TABER BE LIABLE FOR ANY INJURIES DAMAGE TO OR REPLACEMENT OF OTHER PRODUCTS OR PROPERTY COSTS FOR RECOVERING REPROGRAMMING OR REPRODUC ING ANY PROGRAM OR DATA USED WITH THE PRODUCT...

Page 5: ...t the wheels traverse a complete circle on the specimen surface revealing abrasion resistance at all angles relative to the weave or grain of the material It is important to recognize there may be sev...

Page 6: ...t the vacuum unit to the Abraser 6 Plug the female connector of the Abraser power cord into the power inlet located on the rear of the instrument Connect the instrument to a 115V 60 Hz or 230V 50 Hz c...

Page 7: ...button and then select the desired option After making your selection press OK NOTE Options shown below in bold text are factory default settings Language English Mandarin Japanese Spanish French Ger...

Page 8: ...test parameters When the instrument is not operating press the appropriate button and use the dial or odometer to set the value After making your selection press OK Test Cycles The Abraser can be pre...

Page 9: ...shown for the correct specimen holder To change between the left or right press the arrow button on the side of the display screen or press the toggle switch at the top status bar LEFT or RIGHT will...

Page 10: ...uum suction for accurate standardization and control of test procedures WARNING Do not use the vacuum for wet tests To minimize risk of shock disconnect the electrical cord to the vacuum unit NOTE Whe...

Page 11: ...be refreshed at intervals during the test NOTE Do not exceed 50 resurfacing cycles on an S 11 refacing disc If more than 50 cycles are required to refresh the wheels use additional refacing discs Vac...

Page 12: ...Operating Instructions ver 1 0 p n 135505 Display Option The Rotary Platform Abrasion tester can be programmed to show Cycles or Time Both Cycles and Time can be displayed as ascending completed Cycle...

Page 13: ...e way onto the mounting hub Release the push button The hub retaining lip will secure the wheel until it is disengaged Genuine Taber abrading wheels are supplied in pairs A close inspection of the whe...

Page 14: ...al counterweight kit is available when less than 250 g load is required Vacuum System Models 1700 1750 include a Taber vacuum unit that includes a heavy duty motor and life lubricated sealed bearings...

Page 15: ...larly defined intervals Do not confuse coloring of the wheels with clogging as this may simply be due to the wheel faces becoming coated with fine abrading dust a condition that must be expected It is...

Page 16: ...and mount the test specimen Select CONTINUE to continue testing or REFACE to perform a second refacing NOTE The S 11 refacing disc is used only once for a maximum of 50 cycles Never reface wheels wit...

Page 17: ...ter edge of each wheel Manually turn the nut to rotate the motor shaft Reverse the wheels on the arbor and round the opposite corners Only the extreme sharpness of the edges should be removed A radius...

Page 18: ...plate and nut and finally tighten the hold down ring Avoid tightening to excess so as not to wrinkle the specimen Trim off any excess material around the edges WARNING To prevent damage to the instru...

Page 19: ...wear life under changing atmospheric conditions as in automobile and aircraft materials the atmosphere of the laboratory may be controlled accordingly NOTE If an environmentally controlled test labor...

Page 20: ...y 3 Secure specimen to Abraser specimen holder 4 Select and mount the auxiliary weights if necessary 5 Adjust the vacuum nozzle gap 6 Using the Abraser touchscreen enter the appropriate test parameter...

Page 21: ...iable method that records the weight loss of the test specimen due to abrasion This method is recommended when the results are to be compared with those of similar materials having nearly the same spe...

Page 22: ...cific gravities a correction factor must be applied to give a true indication of the wear resistance In doing this the aluminum material evidences a much greater volume loss as seen by comparison of t...

Page 23: ...es or more it is often attributed to procedural errors or an instrument that is out of calibration or one of the factors mentioned below Physical properties of the material such as hardness or resilie...

Page 24: ...for unpaired data at a probability level chosen prior to the testing series If bias is found either its cause must be found and corrected or future test results must be adjusted in consideration of t...

Page 25: ...5505 HELP At the bottom of the display screen is the HELP button Pressing this will provide details on the instrument USB Port Incorporated into the Model 1700 1750 Abraser is a USB port This is used...

Page 26: ...he holder and lift straight upward To install the specimen holder align the set screws with the flats on the motor shaft Cleaning Vacuum Pick Up Nozzle Abraded material and debris may collect in the v...

Page 27: ...14120 USA Phone 716 694 4000 Fax 716 694 1450 E mail sales taberindustries com WebSite www TaberIndustries com www OrderTaber com Part Description 133338 Quick Release Wheel Hub Collet Kit replacement...

Reviews: