1
Pre-operative preparation
Required instruments
Set Pelvic C-Clamp
01.306.000
The Pelvic C-Clamp set consists of the instruments and implants
for the emergency treatment of the indicated fracture types. The
bottom layer of the Vario Case houses the Pelvic C-Clamp and
the optional pliers. The upper layer houses the cannulated nails
in two lengths and the remaining instruments.
The complete sterile Pelvic C-Clamp set should be kept ready for
use in the resuscitation room. Depending on the type of injury,
the orientation points in the pelvic region of the injured person
may be unclear. Should there be doubts about the anatomic
references, use an image intensifier during application of the
Pelvic-C-Clamp.
Pre-operative preparation
– Anteroposterior (AP) plain pelvic radiograph if necessary,
oblique views (Inlet and Outlet) or CT.
– Patient positioning must allow for intraoperative fluoroscopic
controls in AP, Inlet and Outlet Projections.
– Have an image intensifier available.
Positioning
– Place the patient in a supine position.
– To facilitate reduction, ensure free draping of the leg on the
injured side. Strive for good draping coverage of the genital
region.
– To prepare the patient, disinfect the proximal femur and the
buttocks and cover with sterile sheets.
Technique Guide
6
Synthes
Pelvic C-Clamp
Technique Guide