4
Placement of the Pelvic C-Clamp
Required instruments
Pelvic C-Clamp, complete
03.306.010
Nail for Pelvic C-Clamp, cannulated, short, length 190 mm
02.306.006
Nail for Pelvic C-Clamp, cannulated, long, length 210 mm
02.306.007
Ratchet Wrench for Nut, hexagonal, 11 mm
321.200
Wire Cutter, large, with multiplication, length 220 mm
391.930
After inserting the Kirschner Wire on the uninjured side, slide
the clamp with cannulated nails over the wire and ensure that
the tip of the nail grips the bone securely. Then place the second
nail on the injured side (no Kirschner wire is necessary on this
side).
Note:
In cases of severe dislocations of the pelvis, pulling on the
leg, internal rotation and even lateral compression may improve
reduction and facilitate application of the Pelvic C-Clamp.
10
Synthes
Pelvic C-Clamp
Technique Guide