Synthes
21
3
Remove locking bolts
Unscrew the locking bolts using the Hexagonal Screwdriver
(314.750). If the bolts cannot be unscrewed because the
thread no longer grips the bone, use the Holding Sleeve
(314.280).
4
Mount hammer guide
Before removing the last locking implant, screw the Extraction
Screw for DFN (357.133) into the distal end of the nail and
tighten with the Pin Wrench (321.170). The last locking
implant prevents the nail from rotating or sliding away.
Screw the Hammer Guide (357.117) onto the extraction
screw for DFN and tighten.
Alternative
If the extraction screw for DFN is not available, the hammer
guide can be screwed directly into the distal end of the nail.