DC POWER CABLES
PAGE
|
69
DC power cables
•
DC power cables
•
DC voltage sense cable
DC power cables
The DC power cables shipped with p250 lasers are manufactured with 1/0 AWG wire to a standard
length of 2.0 m (6.5 ft) or an optional length of 5.0 m (16 ft). Nominal finished O.D. is 14.9 mm (0.586") so
the minimum bend radius must be greater than 12 cm (4.7 in). Terminals on the laser end of the cables fit
the laser’s M10 studs while terminals on the power supply ends are sized to fit M6 (0.25") bolts.
When using a power supply other than the PS-401, we recommend using remote sensing, so the power
supply output remains at a constant voltage over varying load conditions. Choose a DC supply that can
compensate for a minimum load lead loss (round trip) of 1.0 V.
Important Note: If you lengthen the DC Power Cables shipped with the p250 laser, you must
calculate and measure the additional voltage drop to ensure that 48.0 VDC is available at the
laser’s +48V POWER terminal under full-load conditions. Keep in mind, depending on the additional
length required, you may need to use larger gauge (2/0) wire.
Summary of Contents for p250
Page 1: ...ENGINEERED BY SYNRAD p250 Laser User Manual...
Page 15: ...NOMENCLATURE CONTINUED PAGE 15 Nomenclature Continued Figure 1 2 Anatomy of a model number...
Page 27: ...EUROPEAN UNION EU REQUIREMENTS ROHS COMPLIANCE PAGE 27 Figure 2 3 p250 Declaration Document...
Page 40: ...RF POWER SUPPLY PAGE 40 Table 4 3 Dew point temperatures Continued...
Page 63: ...INPUT OUTPUT SIGNALS PAGE 63 Table 4 4 User I O pin descriptions Continued...
Page 84: ...P250 GENERAL SPECIFICATIONS PAGE 84 Table 4 5 p250 general specifications Continued...
Page 87: ...P250 OUTLINE MOUNTING DRAWINGS PAGE 87 Figure 4 47 p250 packaging instructions...
Page 93: ...OPERATIONAL FLOWCHART PAGE 93 Operational flowchart Figure 5 1 p250 operational flowchart...
Page 94: ...STATUS LED S PAGE 94 Status LED s Table 5 2 p250 Input Output LED Status Signals...
Page 105: ...INDEX PAGE 1 This page is intentionally left blank...