Cooling connections
(cont.)
Cooling tubing connections
The following procedure will guide you in configuring the most efficient cooling system. Please connect your system exactly
as described below.
To connect cooling tubing to your v30 laser, refer to Figure 3 and perform the following steps.
OUTLET
INLET
Chiller
Figure 3
OEM v30 cooling connections
1
Cut and connect a length of tubing to fit between the chiller’s Outlet port and the upper
WATER IN
port on the rear
of the OEM v30 laser.
2
Cut and connect a length of tubing to fit between the lower
WATER OUT
port on the rear of the laser and the
chiller’s Inlet port.
Caution
possible
equipment
damage
Inlet cooling water temperature must always be maintained above the dew point
to prevent condensation and water damage to your Firestar laser.
Caution
possible
equipment
damage
Operating the laser at coolant temperatures above 22 °C (72 °F) may result in
decreased performance and/or premature failure of electronic components.
3
Turn on the chiller and adjust the temperature setpoint to between 18 °C to 22 °C. Regulate coolant flow to 3.8 lpm
(1.0 GPM) at less than 414 kPa (60 PSI) of pressure.
4
Closely examine all cooling connections and verify that there are no leaks.