operation
31
SYNRAD OEM v30 Operator’s Manual Version 2
Controls and indicators
OEM v30 rear panel
Figure 3-2
OEM v30 rear panel controls and indicators
4
DC Power Cables – re30 VDC from the DC power supply. The DC power cables are manu-
factured from #12 AWG wire and measure 1 meter (40 inches) in length. The red (positive) cable
contains a replaceable in-line fuse. If fuse replacement is required, replace it with a Cooper Buss-
mann ABC-20-R, a Littelfuse 314020 or equivalent 20 A, fast-blow fuse.
5
Interface A (DB-9) Connector – provides a connection point for auxiliary output power as well as
input and output signals. See Interface connections in the Technical Reference chapter for DB-9 inter-
face details and pinouts.
6
Interface B (RJ45) Connector – provides a connection point for auxiliary output power as well as
input and output signals. See Interface connections in the Technical Reference chapter for RJ45 inter-
face details and pinouts.
7
LASE Indicator – illuminates red to indicate that the laser is actively lasing. The LASE indicator is
off when tickle pulses are being generated and illuminates red when PWM Command signal pulses
are long enough to produce laser output.
8
RDY (Ready) Indicator – illuminates yellow when the laser is enabled, indicating that lasing will
begin when a PWM Command signal is applied.
9
PWR (Power) Indicator – illuminates green when +30 VDC power is applied to the laser.
10 WATER IN Port (water-cooled models only) – labeled IN, this connection provides the cooling
water inlet to the v30’s coolant path.
11 WATER OUT Port (water-cooled models only) – labeled OUT, this connection provides the cooling
water outlet from the v30’s coolant path.
8
9
6
5
4
INTERFACE
LASE
RDY
PWR
A
B
7
10
11
(Water-Cooled v30 Shown)