getting started
9
SYNRAD OEM v30 Operator’s Manual Version 2
Connecting
Table 1-2
Dew point temperatures
Dew Point Temperature Chart °F (°C)
Relative Humidity (%)
20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95
Air Temp
°F (°C)
60 (16)
— — — 32 36 39 41 44 46 48 50 52 54 55 57 59
(0) (2) (4) (5) (7) (8) (9) (10) (11) (12) (13) (14) (15)
65 (18)
— — 33 37 40 43 46 48 51 53 55 57 59 60 62 64
(1) (3) (4) (6) (8) (9) (11) (12) (13) (14) (15) (16) (17) (18)
70 (21)
— 33 37 41 45 48 51 53 56 58 60 62 64 65 67 69
(1) (3) (5) (7) (9) (11) (12) (13) (14) (16) (17) (18) (18) (19) (21)
75 (24)
— 37 42 46 49 52 55 58 60 62 65 67 68 70 72 73
(3) (6) (8) (9) (11) (13) (14) (16) (17) (18) (19) (20) (21) (22) (23)
80 (27)
35 41 46 50 54 57 60 62 65 67 69 71 73 75 77 78
(2) (5) (8) (10) (12) (14) (16) (17) (18) (19) (21) (22) (23) (24) (25) (26)
85 (29)
40 45 50 54 58 61 64 67 70 72 74 76 78 80 82 83
(4) (7) (10) (12) (14) (16) (18) (19) (21) (22) (23) (24) (26) (27) (28) (28)
90 (32)
44 50 54 59 62 66 69 72 74 77 79 81 83 85 87 88
(7) (10) (12) (15) (17) (19) (21) (22) (23) (25) (26) (27) (28) (29) (31) (31)
95 (35)
48 54 59 63 67 70 73 76 79 81 84 86 88 90 92 93
(9) (12) (15) (17) (19) (21) (23) (24) (26) (27) (29) (30) (31) (32) (33) (34)
100 (38)
52 58 63 68 71 75 78 81 84 86 88 91 93 95 97 98
(11) (14) (17) (20) (22) (24) (26) (27) (29) (30) (31) (33) (34) (35) (36) (37)
To use Table 1-2, look down the Air Temp column and locate an air temperature in Fahrenheit or Cel-
sius (°C values are shown in parentheses) that corresponds to the air temperature in the area where your
laser is operating. Follow this row across until you reach a column matching the relative humidity in your
location. The value at the intersection of the Air Temp and Relative Humidity columns is the Dew Point
Temperature in °F (or °C). The chiller’s temperature setpoint must be set above the dew point tempera-
ture. For example, if the air temperature is 85 °F (29 °C) and the relative humidity is 60%, then the dew
point temperature is 70 °F (21 °C). Adjust the chiller’s temperature setpoint to 72 °F (22 °C) to prevent
condensation from forming inside the laser.
Caution
possible
equipment
damage
Operating the laser at coolant temperatures above 22 °C (72 °F)
may result in decreased laser performance and/or premature failure of
electronic components.