7
E
Footswitch
Comment: It is also possible of powering the D62S display via the mini-USB, in case of the
periphery is capable of giving 5V/1000mA.
Pin 1 : Ground
Pin 2 : Input +9.0V (not used for footswitch)
Pin 3 : External contact input 1 (signal = 0 Volt)
Mini-USB connector
The mini-USB connector has two functions: HID or Virtual port COM
1. HID: transferring measurements only. If you connect the D62S to a PC, it will be
detected as a keyboard using the standard drivers for the operating system. When
you transfer a measurement, the number will be «written» to the PC as if you had
typed it in on a keyboard.
2. Virtual Port Com: this connection has the same parameters than the RS232 Port:
4800, E, 7, 2
(or according to user configuration)
It can be used either for transferring measurements or sending commands from
the computer. Measurement can be sent to any windows application or software
using Sylvac Vmux application or Sylcom. To use the USB output as a virtual
com port, you will need to install specific drivers (available on www.sylvac.ch for
downloading). Windows message confirming that the D62S0 has been correctly
detected and installed: