34
The D62S internal software can be updated in case of necessary.
Updating requires an RS232 1:1 cable. It is possible to use an RS232/USB converter if
your PC is not equipped with a serial port.
Updating the firmware requires the «flash magic» software, which can be downloaded free
of charge from this address: www.flashmagictools.com.
After installation, configure the software as follows :
16. Firmware update
Procedure :
1. Connect the D62S to the computer with standard RS232 cable 1:1
2. Connect the footswitch
3. Switch on the D62S while pressing footswitch (screen will remain white)*
4. Configure the flash magic software according to the image above.
5. Click on “Start”. Updating will take between 2 and 5 minutes depending on the speed
chosen. During updating the screen is locked.
6. The D62S restarts automatically when the procedure is completed.
7. Reinitialise the D62S.
*Comment: check on which Com Port port your unit is connected, and select corresponding
port on COM Port