26
Selection of the end of line.
The D62S uses the Sylvac communications protocol in order to run and configure its functions.
Data transmission format
Data transmission can be requested either by favourite button, footswitch or request
from computer.
Comment : request or command is made on the active channel.
2-channel mode:
D62S: the values of the 2 channels are transmitted, separated by a comma «,»
D62S ADV: The values of the 2 channels are transmitted, separated by a comma «,» or of
the active channel if one of the 2 channels has been manually selected (refer to Chapter
Printing for available configurations).
Transmission format for 1 channel: +/- E0···En DP D0···Dn CR
Example: +123.456<CR>
Transmission format for 2 channels: +/- E1···E1n DP D10···D1n , +/- E20···E2n DP D20···Dn CR
Example: +123.456,+234.567<CR>
13. Communication protocol
List of commands
Command
Reply
Function
D62 Adv.
?
+|-[xx]xx.yy[yyy][_< | = | > ]
Data request: active channel
?x
data request of channel (x)
Bat ?
BAT0/1
Battery status
BEEP ?
BEEP0/1
Status buzzer
BEEP 0
Deactivate the buzzer