SYCLOPE electronic TRACE'O Installation And Starting Instructions Download Page 31

Starting instructions 

 

Page 31/40 

 Installation and starting instruction manual 

SYCLOPE TRACE’O

®

 

 

2)

 

Setting real-time clock 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

          Adjust the real time clock 

 

 

 

 

 

 

 

 

 

 

 

3)

 

Contrast adjust (Only when « Stopped ») 

From the main screen (displaying the measurements), hold the "+" or "-" key pressed to adjust the 

display contrast. 

 

 

  High 

 

 

 

 

 

 Low 

 

4)

 

Setting display screens 

According to the measured parameter of the 

SYCLOPE TRACE’O

®

 controller, it is possible to change 

the manner for displaying the values on the main screen.  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  

Setting real time  

_

  

 

 Year:15   Month:02 

 

 Date:19 

 Day:04 

 

Time.....:17H14min 

>>   MAIN MENU   <<

_

  

User menus 

Langue/Sprache/Taal 

Setting real time 

Services 

 

 *   TRACEO V1.00   * 

  Sn: xxxxxxxxx 

 

>>   MAIN MENU   <<

_

  

User menus 

Langue/Sprache/Taal 

Setting real time 

Services 

 

 *   TRACEO V1.00   * 

  Sn: xxxxxxxxx 

 

>>   USER MENU   <<

_

  

Specialist menus 

User code 

Calibrations 

Setpoints 

Timers of cycles 

Technical alarms 

Analog. outputs 

 

 

 

 

>>   MAIN MENU   <<

_

  

User menus 

Langue/Sprache/Taal 

Setting real time 

Services 

 

 *   TRACEO V1.00   * 

  Sn: xxxxxxxxx 

 

>>   USER MENU   <<

_

  

Calibrations 

Setpoints 

Timers of cycles 

Technical alarms 

Analog. outputs 

Printer output 

Display management 

>>   USER MENU   <<

_

  

Display management 

Back-light 

Main screen 

Last line 

K:- - - -                                      

   

14:44 

 

    ->.<- : 1.50                   ->.<- : 7.40 

 

1.50 

Cl 

pH 

7.86 

Summary of Contents for TRACE'O

Page 1: ...Installation and starting instructions Reference CTR 0001 Rev 2 SYCLOPE TRACE O Universal controller for trace detection Part 1 ...

Page 2: ...ion SYCLOPE Electronique 2016 Manual of 2016 February 18th Rev 2 Universal controller for standard and trace measurement Product line TRACE O Part 1 Installation and starting instructions Ref DOC0321 Editor SYCLOPE Electronique S A S Z I Aéropole pyrénées Rue du Bruscos 64 230 SAUVAGNON France Tel 33 05 59 33 70 36 Fax 33 05 59 33 70 37 Email syclope syclope fr Internet http www syclope fr 2015 20...

Page 3: ...ee of potential relays KA KB KC et KD 17 6 Connection of the measurement inputs 17 7 Connections of the analogue outputs IA1 and IA2 20 8 Connection of the remote control input CAD 20 9 Connection of the flow control entry sensor Flsw or low level detection tank 21 10 Connection of the counter entry Flow or low level detection tank 22 11 Connection of the RS232C printer output 22 12 Branchements d...

Page 4: ...s TRACOM software also developed by SYCLOPE Electronique S A S makes it possible to ensure the maintenance and the programming of the basic functions as well as the functions necessary to the communications with the Internet With an adequate interface carrying out internet connectivity SYCLOPE TRACE O will be connected in real time to Internet site mysyclope com thus ensuring a management and a to...

Page 5: ...uctions Information Action to be taken Item of a list or catalogue 3 Symbols and signs Identification of a continue voltage or current Identification of an alternative voltage or current Protective ground Functional ground Risk of injury or accident Identify a warning concerning a potentially dangerous risk Documentation must be consulted by the user with each time the symbol is notified If the in...

Page 6: ...with according electrical glands in compliance with IP65 level protection Cables must be adapted to the electrical glands to respect the level of protection Grounded cables for connecting pH and ORP Redox sensors are not provided Content of the packaging One analyzer controller SYCLOPE TRACE O Installation and starting instruction notice Programming notice Communication notice Option 6 Warranty Th...

Page 7: ...kes not to allow its employees to work with the SYCLOPE TRACE O equipment described in this manual unless they Are aware of the fundamental instructions relating to work safety and prevention of accidents Are trained in the use of the device and its environment Have read and understood these instructions warnings and manipulation rules 3 Risk prevention The installation and connection of the SYCLO...

Page 8: ...njections can be noted In the doubt a service contract must be taken near your reseller installer or failing this near our engineering services Contact your approved installer reseller or our business service for more information 4 Identification and localization of the identification plate Label of the manufacturer Serial number Model of the product Particular risks Read de manual Reference of th...

Page 9: ...ot be thrown out together with household or industrial waste According to current regulations consumers within the European Union are required as of this date to return their used devices to the manufacturer who will take care of disposing them at no extra expense According to European directive 2011 65 UE this symbol means that the SYCLOPE TRACE O controller is designed in compliance with the res...

Page 10: ...1 2 kg Protection rating IP 65 Display Blue back light 128x64 LCD screen Inputs Measurement inputs 2x 4 20mA inputs on special configuration modules Auxiliary input 1x 4 20mA auxiliary input for additional function Option Control inputs 1x control input for external Start Stop 1x control input for level tank or flow level input CAD Flsw Metering input 1x flow meter pulse input or chemical product ...

Page 11: ...t values Control of top and bottom thresholds Closed loop control Remote control Flow rate control Closed loop control of injections with an external contact filtering for example or with flow switch control from cell Level tanks Low level detection for chemical tanks Control of each injection channel to one or two external contact s Flow et Flsw Timers Programming of polarisation depolarisation c...

Page 12: ...g l 0 to 2mg l 0 to 5mg l 0 to 10mg l 0 5 Active chlorine 0 to 1mg l 0 to 2mg l 0 to 5mg l 0 to 10mg l 0 to 20mg l 0 to 50mg l 0 to 100mg l 0 to 200mg l 0 to 250mg l 0 5 Total chlorine Trace function 0 to 1mg l 0 to 2mg l 0 to 5mg l 0 to 10mg l 0 to 20mg l 0 5 Chlorine dioxide 0 to 0 5mg l 0 to 1mg l 0 to 2mg l 0 to 5mg l 0 to 10mg l 0 5 Chlorites 0 to 1mg l 0 to 2mg l 0 5 Bromine 0 to 1mg l 0 to ...

Page 13: ... l 0 to 10mg l 0 to 20mg l 0 to 50mg l 0 to 100mg l 0 to 200mg l 0 to 500mg l 0 to 1000mg l 0 to 2000mg l 0 to 5000mg l 0 to 1000mg l 0 5 PHMB 0 to 50mg l 0 to 100mg l 0 to 200mg l 0 to 500mg l 0 5 Input current 0 to 20mA 0 5 b Version with 1 isolated potentiometric entry and 1 isolated 4 20mA entry The main entry can be programmed according to the version as bellow Measurements and controls of th...

Page 14: ... a c Version with 2 potentiometric entries The main entry and the auxiliary entry can be programmed only as bellow Measurements and controls Parameter Range of measurement Accuracy pH 2 to 11 pH 0 5 Rédox 0 to 1000 mV 0 5 Rédox 0 to 1500 mV 0 5 Rédox 1000 mV 0 5 Rédox 1500 mV 0 5 SYCLOPE TRACE O controller will detect automatically the type of the additional module ...

Page 15: ...e warranty is not applicable 2 Installation of the wall mounted controllers Before performing the installation and electrical connections remember to turn off the power The rating of IP65 is only guaranteed when the closing cover and the glass of the electric box are closed and when the cable glands match the diameters of your cables and are correctly sealed Drilling three holes 5 mm according to ...

Page 16: ...er current and by a varistor against voltage surges over 275 V Reference Name FUS5X20T315 Time lag fuse 315mA 5x20 Glass In case of fuse destroyed check that the card is not burnt If this is the case the complete card must be changed In case of varistor burned please return the controller to our technical after sales department for repairing 4 Connection of the primary power supply The SYCLOPE TRA...

Page 17: ...controller is equipped with three entries maximum according to the module configuration Using SYCLOPE cells is mandatory They are adapted to the different type of sensors and offer the best guaranty of good measurement No claim will be received in case of unrespect of this condition a Basic version 2 entries 4 20mA NOT isolated The amperometric measurement inputs generate their own power supply an...

Page 18: ... terminal Entry White wire Right terminal c Potentiometric and mA entries module 1x mV 2x 4 20mA The amperometric measurement inputs E1 and E2 generate their own power supply and should not be powered in any case The main potentiometric entry E1 is isolated E1 The auxiliary 4 20mA entry E2 is isolated E2 The last passive entry is not isolated E3 Please respect the polarity of the amperometric sens...

Page 19: ...f the shielded cable e Connecting a 4 20mA Active sensor E1 and E2 entries cannot be used with an active sensor when isolated Normally the E1 and E2 entries are powered and E3 entry is passive No power generated When connecting an active sensor power generated by another source the wiring must be executed as bellow When using an active sensor of the active output sensor to the entry E1 E2 or E3 of...

Page 20: ...u can therefore assign an output to any measured parameter and use it for regulation or transfer operations 8 Connection of the remote control input CAD The SYCLOPE ODISEA controller has a remote control input CAD which stops the regulation units This input is an open closed contact input used in a subservient manner to the main circulation pump of the pool filtration system It is compulsory to sl...

Page 21: ... control function this input can be also used to control the low level of chemical product tanks and will cause the stop of the dosage of the corresponding products when the low level is reached In this last case the function NO Normally open or NC Normally closed is programmable a Case of a connection to a potential free dry contacts 2wires The detector must be free of potential contacts Referenc...

Page 22: ... the function NO Normally open or NC Normally closed is programmable 11 Connection of the RS232C printer output The SYCLOPE TRACE O controller has a serial compatible RS232C output speed 4800 bauds for printing paper reports guaranteeing surveillance of your measurements and editing the operating log of the machine SYCLOPE Electronique S A S proposes you a compatible printer like that a connecting...

Page 23: ...nterface module to connect your SYCLOPE TRACE O controller to your computer Please consult the instructions of this controller for the connection Réference Name INF1021 USB 485 converter 13 Connections of the MODEM port for communications The SYCLOPE TRACE O controller has a RJ45 Modem output for connection to a phone line to establish a remote link with a computer via the SYSCOM or TRACOM communi...

Page 24: ... to the type of modem and subscription of internet connectivity the data are transmitted to the data web site mysyclope com allowing thus a management in real time of the swimming pool Messages of alarms can be sent to the users by emails or SMS and a history of measurements and alarms is recorded GSM modem kit WIFI modem kit Ethernet modem kit Reference Name Contact us for quotation 15 Connection...

Page 25: ...ows you to connect all the entries and outputs of the SYCLOPE TRACE O controller The quantity of glands of the SYCLOPE TRACE O controller is only sufficient for connecting simple needs but use more glands in case of using full functions This extension kit is deliver with a multi wires special cable to be connected quietly to the SYCLOPE TRACE O controller ...

Page 26: ... sampled in a specific sensor housing located after the recirculating pump This sensor housing receives the water to be analyzed and transmits all parameters of the measurement probes to the SYCLOPE TRACE O controller According to the setting points programmed by the user the SYCLOPE TRACE O controller sends to the dosing units all informations of chemical product dosage and injects them after the...

Page 27: ...e controlled is sampled from the measurement circuit Water for checking the sensor is sampled from the polarization circuit Measuring cell receives the sensor for measuring chemical product and sends the value to the controller Beginning of the polarisation cycle Polarisation step o Solenoid valve A closed o Solenoid valve B opened o Waiting to reach the programmed threshold Starting the polarizat...

Page 28: ...pad Backlit 64x128 display with white writing on blue background Menu key To allow accessing to the programming menu red LED Calibration key To allow the sensors to be directly calibrated Orange Led STOP START key To allow starting or stopping of the processes green LED Esc key To allow to delete the settings or to come back in the programming menus OK key To allow to validate the settings or to m...

Page 29: ...s F1 General fuse Time lag 315 mA miniature Glass type Terminal block See below Modem phone line standard connector RTC Location for socket modem Phone GSM WIFI or Ethernet optional Serial printer connector RS232C Ribbon cable connector to the top card 3 Terminal block connections 1 2 F1 1 2 3 4 5 6 3 4 5 5 ...

Page 30: ...of your installation are not disrupted The SYCLOPE TRACE O regulator does not automatically begin treatment and dosing of chemical products when switched on Only the user can control when to begin treatment having checked that the central unit has been correctly programmed according to his her needs When switching on the measured parameters predefined by the basic configuration are displayed and t...

Page 31: ...eal time _ Year 15 Month 02 Date 19 Day 04 Time 17H14min _ MAIN MENU _ User menus Langue Sprache Taal Setting real time Services TRACEO V1 00 Sn xxxxxxxxx _ MAIN MENU _ User menus Langue Sprache Taal Setting real time Services TRACEO V1 00 Sn xxxxxxxxx _ USER MENU _ Specialist menus User code Calibrations Setpoints Timers of cycles Technical alarms Analog outputs _ MAIN MENU _ User menus Langue Sp...

Page 32: ...n below b Case of two parameters on screens According to the selected parameters display the general aspect is shown below _ USER MENU _ Display management Back light Main screen Last line _ USER MENU _ Main screen Channel E1 Cl Channel E2 pH Channel Flow rate Two screens Three screens K 14 44 7 40 7 56 pH _ USER MENU _ Main screen Channel E1 Cl Channel E2 pH Channel Flow rate Two screens Three sc...

Page 33: ...nvironment Adjust the value of the setpoint Repeat this process for all the other setting items _ USER MENU _ Main screen Channel E1 Cl Channel E2 pH Channel Flow rate Two screens Three screens _ USER MENU _ Specialist menus User code Calibrations Setpoints Timers of cycles Technical alarms Analog outputs _ MAIN MENU _ User menus Langue Sprache Taal Setting real time Services TRACEO V1 00 Sn xxxxx...

Page 34: ...ng the operations to perform please contact our technical department before calibration An incorrect calibration can result in excessive doses of the chemical product and thus harm the environment The SYCLOPE TRACE O controller has a key on its front panel allowing you to perform the calibration directly This direct key is operational while the treatment is active green STOP START key lit _ USER M...

Page 35: ...s the treatments as required The LED of the STOP START button can have various statuses according to the environment Lit the processes are active Not lit the processes are inactive Flashing the processes are paused if a start delay is programmed or if the conditional optional functions CAD and Flsw are active or if an operating interval timer has been defined and when the controller is out of this...

Page 36: ...ammed as a low level tank Notification about flow rate Flow when programmed as a low level tank Indicate that internet connection to www mysyclope com is active General notification area K CAD FLS 14 44 7 40 7 01 pH FLO General notification area Channel notification area Information area Last line Trace detection notification area Cl K A CAD 14 44 7 40 1 5 7 01 pH FLO Information area Last line FL...

Page 37: ...e high threshold activated Indicate a technical alarm Injection of chemistry activated Open phase of 3 ways command Close phase of 3 ways command Indicate Pause of all treatments When tank is empty or flow switch entry is actived Signs of Last line area Setpoint of the programmed channel IX Guide and maintenance The controller does not require any specific maintenance Repairs may only be performed...

Page 38: ...CE Certificate Page 38 40 Installation and starting instruction manual SYCLOPE TRACE O ...

Page 39: ...Installation and starting instruction manual SYCLOPE TRACE O ...

Page 40: ...tronique S A S Z I Aéropole pyrénées Rue du Bruscos 64 230 SAUVAGNON France Tel 33 05 59 33 70 36 Fax 33 05 59 33 70 37 Email syclope syclope fr Internet http www syclope fr 2015 2016 by SYCLOPE Electronique S A S ...

Reviews: