Commissioning
Page 18/40
Installation and starting instruction manual
SYCLOPE TRACE’O
®
b)
Full mA entries module (3x entries 4…20mA)
The amperometric measurement inputs generate their own power supply and should not be
powered in any case!
The main and auxiliary entries E1 and E2 are isolated (E1 & E2).
The last passive entry is not isolated (E3).
Please, respect the polarity of the amperometric sensor!
Entry
+ : Blue wire (Left terminal)
Entry
- : White wire (Right terminal)
c)
Potentiometric and mA entries module (1x mV & 2x 4…20mA)
The amperometric measurement inputs (E1 and E2) generate their own power supply and
should not be powered in any case!
The main potentiometric entry E1 is isolated (E1).
The auxiliary 4…20mA entry E2 is isolated (E2).
The last passive entry is not isolated (E3).
Please, respect the polarity of the amperometric sensor!
Entry
+ : Blue wire (Left terminal)
Entry
- : White wire (Right terminal)
Temperature probe
without polarity
Amperometric sensor :
Respect polarity
pH or REDOX sensor
Hot point to the left
Amperometric sensor :
Respect polarity