Operating Instructions
21
4.4 Calibration
With and select according to your
significance, with leave the menu
without any change, with transfer the
entry and switch to a deeper menu level.
4.4.1 Calibration Point 1 - Measuring Value
Measuring value in phys. units.
Range: Measuring start - Measuring end
With set the value to 0.0 to start enter the
numbers of the measuring range, with
transfer the entry and leave the menu level.
4.4.2 Calibration Point 1 - Raw Value
Indicate the initial value to the value
displayed, if pressed . With leave the
menu without any change.
All other points are calibrated as the first one.
4.4.3 Calibration Point 2 - Measuring Value
Measuring value in phys. units.
Range: Measuring start - Measuring end
4.4.4 Calibration Point 2 - Raw Value
Indicate the initial value to the value displayed.
4.4.5 Calibration Point N - Measuring Value
Only necessary, if you are disturbed by
the non-linearity
(see diagram on the right side).
The measuring points of the nominal
characteristic curve will be typed in
and calibrated to the actual
characteristic curve. The customization
will be done by the evaluation unit.
The output values are linear.
Measuring value in phys. units.
Range:
Measuring start - Measuring end
4.4.6 Calibration Point N - Raw Value
Indicate the initial value to the value displayed.
C
8
C
8
£
¤
8
C
Calibration
7 8 9
Calibration Point 1
4 5 6
0.0
kg/h
1 2 3
C
0
8
Calibration
Calibration Point 1
0.015663
C
Akt.: 0.015009
8
4.4 Calibration
£
4.4.1 Val. P1
1. 0.00
¤
4.4.2 Calibration P1
4.4.1 Val P2 1.58
C
4.4.2 Calibration P2
8
q
Stützpunktkennlinie
0
10
20
30
40
50
60
70
80
90
100
0
1
2
3
4
5
6
7
8
9 10
Messpunkte
W
er
te
ist
soll
Diff 1
Diff 2
Diff 3
Diff 4
Diff 5
Diff 6
Diff 7
Diff 8
Diff 9
Values
Set Points
Measuring Points