Operating instructions
fore resetting the system by completing the following
steps:
Press button
(17)
on the switching power supply
(15)
until all of the motors have powered down and this
is confirmed by the motor’s emergency light blinking
twice (can take 20 seconds).
The internal meter readings for the individual motors
have now been reset to ‘0’ and the sync function can
work correctly again.
Use of the emergency lowering button consumes a
great deal of battery power, which means you will need
to replace the batteries.
The _75RF / _75RF bridge® does not work despite
being switched on and all components being con-
nected properly
Disconnect the mains supply unit from the mains out-
let. Wait a minute and then reconnect it. If necessary,
connect a bedside lamp to check that mains voltage
(240V) is present. If there is no current, check the fuse.
Press the “all motors down” button
(6)
to reset all ad-
justments. The control of the _75RF / _75RF bridge®
contains microprocessors, so irregular power supply
can lead to malfunctions.
Technical data
Supply voltage:
110-240 V
Remote control power supply:
3 x AAA Batteries
Power rating motors:
Intermittent operation from 2 min / 18 min
Power rating user socket :
2000 W / 10 A
Thrust motors:
2 x 6000 N
Power supply for emergency lowering:
2 x 9 V block battery
Problem
Possible cause
Solution
Remote control
or control system
not working.
No mains supply.
Restore mains connection.
Manual switch or control system
defective.
Contact your supplier / stockist.
Control systems
suddenly no
longer travel /
move.
Thermoswitch on the transformer
may have been triggered or may be
defective.
Leave the control system in the rest position for 20-30
minutes.
Thermal cut-out in the transformer
may have been triggered or may be
defective.
Contact your supplier / stockist.
Internal fuse may have been triggered
or may be defective.
Contact your supplier / stockist.
No mains supply.
Restore mains connection.
Supply line (mains and / or manual
switch / auxiliary control system)
interrupted.
Check the supply line and restore contact if applicable.
The battery-oper-
ated electric reset
function cannot
be activated.
Block battery / block batteries flat.
Check and replace the block battery / block batteries if
necessary.
Block battery/batteries not connected.
Connect the block battery / block batteries.
The radio remote control does not work (The back-
ground light does not work)
Insert new batteries. Where applicable, remove any
obstacles between the radio remote control and the
motor control.
After a power failure once power has been restored
to the system
To restart the initial setup of the uni 14_/15_/22_/20
_75RF / _75RF bridge®RF / _75RF / _75RF bridge®RF
bridge®, press button
(17)
on the switching power
supply
(15)
until all of the motors have powered down
and this is confirmed by the motor’s emergency light
blinking twice. (Can take 20 seconds).