OMEGA Scan'O'Vision MYRIA
Version 1.0
3503.502.02
Page 9
Select the "Advanced" tab; select "
Link Speed/Duplex Mode
" in the list and change the
value from "
Auto Negotiation
" to "
1000Mbps/Full Duplex
". Note: the previous text in
magenta
may change depending of the network adapter.
Close all the opened windows by clicking on the corresponding "OK" button.
3.4
Software use
You can read OSV-STAR and OSV MYRIA images with any OSV7 software configuration.
The complete use of the software is described in the manual 3434.501 (available in the
\Documentation sub-directory. Following are some change related to MYRIA camera.
3.4.1
Information about cameraWith "Help", "About Hardware" menu, you can get a lot of
information about the camera; for example its inclination or the temperature inside the camera
(hottest point).
3.4.2
Test Window
With OSV MYRIA camera, when ready is set, it is automatically reset if a start pulse is received
(with OSV-STAR camera, Ready is independent of Start impulse).