6.  If  you  want  to  power  the  station  from  the  Power 

supply  connect  the  power  supply  cable  to  the  DC 
Supply socket in the same way. 

 

7.  Press 

<Alt>+<Start>

 on SVAN 979 to turn on the 

system. 

 

8.  Remove  the  protection  cap  from  the  microphone 

and perform a calibration check. 

 

9.  Take  out  from  the  case  with  accessories  the  anti-

bird spike with the foam from the accessories case, 
push the foam upwards and mount the tube on the 
thread.  Then  slide  the  foam  down  so  it  hides  the 
metal parts of the kit. 

  

Summary of Contents for SV 279 PRO

Page 1: ...INSTRUMENTATION FOR SOUND AND VIBRATION MEASUREMENTS SVANTEK Sp z o o Warsaw Poland Phone fax 22 839 00 31 e mail office svantek com pl SV 279 PRO Monitoring Station User Manual Warsaw February 2017...

Page 2: ...ty of any kind either expressed or implied including but not limited to its particular purpose Svantek reserves the right to make improvements and or changes to this manual or to the products and or t...

Page 3: ...SB 272 External rechargeable battery 18 2 2 3 SB 271 Solar panel 19 2 2 4 SP 274 meteo module 19 3 OPERATING THE STATION 21 3 1 Powering 21 3 2 Controller interface 21 3 3 Modes of station operation 2...

Page 4: ...t detector time constants e g Fast Slow Impulse The SA 279 outdoor kit protects the SVAN 979 preamplifier and microphone from weather conditions The SA 279 is made of lightweight materials and can be...

Page 5: ...es a waterproof CHARGER that is designed for an outdoor use Military standard CONNECTORS provide reliable robust and waterproof cable connections ACCESSORIES INCLUDED SVAN 979 Class 1 Sound Vibration...

Page 6: ...d level meter is installed inside but can be removed and used as a hand held meter All accessories fit conveniently into a second carrying case The SV 279 PRO station includes outdoor microphone kit 1...

Page 7: ...t but in many applications essential for reliable system operation and task performance are 1 acoustic calibrator SV 33A see chapter 2 2 1 2 external battery 33 Ah including indoor charger SB 272 see...

Page 8: ...ed as a hand held sound meter To do this the user should 1 switch off the instrument by pressing together Alt and Start Stop push buttons 2 unscrew the input connector and disconnect the input cable 3...

Page 9: ...be absorbed by the silica gel bags optional accessory Note The valve must be closed when the station is used outdoors otherwise the case is not sealed against moisture The valve should be released if...

Page 10: ...e risk of leakage The battery has a pressure relief valves that allows safe dispersal of any excess pressure inside the cell VRLA Battery capacity may vary depending on the ambient temperature The saf...

Page 11: ...ystem examining the communication state and condition of every module of the system and the immediate indication of all problems on the control panel by means of several LEDs or remotely using 3G Inte...

Page 12: ...ted PC The BAT 1 LED will start flashing a green colour 4 After Success is registered on the PC disconnect the SC 56 cable Note In case of any problems with the upgrade switch off SVAN 979 and wait un...

Page 13: ...he PC The measurement results can be analysed later with the use of the SvanPC software SVAN 979 can be easily removed from the case and used as an independent hand held sound or vibration level meter...

Page 14: ...oring station case To extract the batteries the user should switch off the instrument unscrew the bolt take off the black bottom cover of the instrument and slide the battery tubes out The user should...

Page 15: ...d or airport the acoustic signal is perpendicular to the microphone s grid The frequency characteristic of the designed filters is given in App C to the SVAN 979 manual 2 Network should be set to GPRS...

Page 16: ...upply with AC DC converter SB 270 is waterproof Single Output Switching Power Supply which is characterised by Universal AC input Full range 90 264VAC Protections Short circuit Over load Over voltage...

Page 17: ...fic mountings Technical data such as direct and frequency characteristics associated with the microphone are included in the SVAN 979 user manual As an option the user may use desiccator Silikogel The...

Page 18: ...p to three times longer operating time of the monitoring station in comparison to the station s internal 17 Ah battery The SB 272 set includes the SB 273 indoor charger and a cable for connection betw...

Page 19: ...toring station It measures the 6 most essential weather parameters barometric pressure humidity precipitation temperature wind speed and direction and also rain and hail intensity It has low power con...

Page 20: ...On in the Logger Results list path Menu Measurement Logging Logger Results SVAN 979 also integrates weather parameters for the period defined by the parameter Integration Period in SVAN 979 path Menu...

Page 21: ...ues to be powered from the internal battery The status of powering is indicated at the controller s panel by a combination of DC CHARGING BAT 1 or BAT 2 LEDs see Table below Note It is recommended tha...

Page 22: ...f 0 loaded or disconnected when charging colour is changed red orange green green when charged off when charging colour is changed red orange green green when charged BAT 2 off off green if loaded 50...

Page 23: ...rgy Battery charging mode When the SVAN 979 is switched off but there is a power supply voltage from SP 270 or current from solar panel only DC CHARGING and BAT 1 LEDs of the controller will be active...

Page 24: ...f necessary use a tool e g pen to push the SIM fully in 4 Press Alt Start on SVAN 979 to turn on the system The station is programmed to automatically establish a 3G connection with SvanNET It takes u...

Page 25: ...se with accessories and take out the SA 279 microphone outdoor kit 3 Mount the SA 279 on a tripod not included or other stable support 4 Remove the protection cap from the input socket on the case by...

Page 26: ...way 7 Press Alt Start on SVAN 979 to turn on the system 8 Remove the protection cap from the microphone and perform a calibration check 9 Take out from the case with accessories the anti bird spike wi...

Page 27: ...irectly after turning on the controller This procedure includes checking the connections with the instrument the 3G modem optionally also the SP 274 meteo module Additionally system power and internal...

Page 28: ...ibed in chapters 6 8 have been made you can start working with the station via the remote control system provided by SVANTEK The easiest way to establish remote control is to create a user account for...

Page 29: ...ing automatic web publication etc Note Establishing 3G connection requires usage of a SIM card with no PIN protection with activated Internet access Installation of the SIM card is described in Chapte...

Page 30: ...Instrument Wireless Transfer Modem Connection 5 Successful connection to SvanNET is indicated by an icon on a SVAN 979 display S means that connection with SvanNET is established Bars and arrows info...

Page 31: ...Client or TCP Server The TCP Client is a mode of main communication channel in which the 3G Modem is configured to initiate connection to a designated address remote host Each time a reconnection per...

Page 32: ...allows to translate a static name to the internet address The special controlling application uses this name to identify the SV 279 PRO in the Internet as an internet address The properly configured s...

Page 33: ...to it Note SvanNET provides also an e mail service The use of SvanNET simplifies the configuration of the station by only switching on the SvanMail position in the SVAN 979 configuration menu path Me...

Page 34: ...described communication channels also via SvanNET For this purpose SvanPC should be completed with special Remote Communication module which usually is delivered as an option to SvanPC SvanPC with Rem...

Page 35: ...is inside of the monitoring station case Note 5 The correct connection of the microphone is not signalled by the controller therefore it is recommended to perform a test measurement each time the stat...

Page 36: ...50 C 6 Storage temperature range SM 279 PRO and SA 250 40 C to 70 C 9 Working relative humidity range 0 100 RH Power Supply 20 Internal battery Europower EP 17 12 12 V 17 Ah 21 SV 279 PRO power consum...

Page 37: ...nge of 30W to 120 W without control system Note Size and power of the panel depend on the climate of the area where the station operates Measurement Line 30 SVAN 979 Meter analyser Datasheets are avai...

Page 38: ...date Temperature Pressure Humidity Wind 1 s Precipitation 10 s Note SVAN 979 acquire every 1s above data from the SP 274 and average them to the step of Summary Results or Logger see SVAN 979 manual S...

Page 39: ...documents 1 SVAN 979 User Manual www svantek com 2 SvanNET Quick Start Guide www svantek com 3 SvanPC User Manual www svantek com 4 SA 279 Assembly Guide www svantek com 5 GeMalto EHS6 Terminal User...