![Svantek SV 279 PRO User Manual Download Page 15](http://html1.mh-extra.com/html/svantek/sv-279-pro/sv-279-pro_user-manual_1409268015.webp)
Monitoring Station SV 279 PRO
15
1. The
Compensation Filter
should be set to
SA279 (90
o
)
or
SA279 (0
o
)
filter (
path:
<Menu>
/
Measurement
/
Compensation Filter
).
Both
filters
are
dedicated for the permanent outdoor
monitoring application. The characteristics of
the outdoor filters depend on the application:
environmental (the acoustic signal is parallel
to the microphone’s grid) or airport (the
acoustic signal is perpendicular to the
microph
one’s
grid).
The
frequency
characteristic of the designed filters is given
in App. C to the SVAN 979 manual.
2. Network
should be set to
GPRS
(
path:
<Menu> / Instrument / Wireless Connection /
Network
).
3. RS232
should be switched on in the
Communication Ports
window
(
path:
<Menu>
/
Instrument
/
Communication Ports
).
Other settings depend on measurement and transmission type.
All measurement, instrument and transmission settings can be set up via SVAN 979
user interface, or remotely via the server SvanNET or + program.