SV 104A USER'S MANUAL_______________________________________________________
76
7.5
Changing the microphone and windshield
To change the microphone, take off the windshield first (see chapter 3.3). Now, you need to unfasten
the microphone from the mounting head. Unscrew the microphone in a counter-clockwise direction.
To attach a new microphone, screw the microphone on clockwise until it fits tightly.
It is critical to make
sure the sensor is tightly fitted
. Use the NM104I1Z wrench but be gently and be aware not to break or strip
the thread.
Note: Note, when the microphone is changed, the new microphone serial number is
automatically stored into the SV 104A internal memory.
In case the windshield is destroyed/lost by any accident, or it gets too dirty to afford the microphone the
best protection then
SVANTEK
offers
SA 122A_3
three pcs per pack (windscreens for
SV 104A
dosimeter).
7.6
Firmware update
SVANTEK
is committed to continuous innovation path of development, and as such reserves the right
to provide firmware enhancements based on user’s feedback.
To update the instrument firmware:
•
Unpack the provided firmware package (provided as a suitable compressed file).
•
Make sure the unit is turned off.
•
Hold down the
<PROFILE>
key and press the
<ENTER>
key to turn on the unit
.
This
ensures the unit will switch on and enter the special reprogramming
BOOTLOADER
mode.
Figure 7-1
Bootloader update mode view
•
Then connect the USB cable. The
<USB>
text will now appear on the instrument display.
•
Run the
start.bat
file.
•
Successful firmware update will be indicated by relevant message.
•
Turn off the unit.
Note: With use of Supervisor software it is very easy to check if there are
any new firmware releases available for download (see Figure 6-3 commands).