The 19” HVPS-HC Module R012 is specified for the operation with an external reference
voltage. Please check the specification sheet of your specific module for the maximum
allowed reference voltage.
All 19” HVPS-HC Module R012 with a maximum output voltage <= 5kV come with SHV
connectors, independent on their maximum floating voltage.
Therefore it must be taken care off, that the sum of the module output voltage and the
applied floating voltage is not exceeding 5kV in total.
In cases that no reference voltage is applied to the device, the termination plug must
be used to ground the reference input of the 19” HVPS-HC Module R012. With a missing
reference potential the device is not providing any output voltage.
Finish the complete cabling of the 19” HVPS-HC Module R012 before switching on the 19”
Basic Unit and switch off the device first before performing any changes to the cabling.
• Install the 19” HVPS-HC Module R012 into a free slot of the 19” Basic Unit (if not already installed).
• Use appropriate cables (SHV or LEMO 1S) to connect the outputs with the corresponding polarity
suitable for your application.
• Use the termination plugs (SHV or LEMO 1S) to terminate the output with the opposite polarity to GND
(e.g. terminate the negative output of channel 1 in case that the positive output of channel 1 should
be used).
• Alternatively an external reference potential can be applied instead of using the termination plug for
a termination to GND.