- 35 -
5.4 REPLACE LENS FEELER
> Check the lens scanner regularly
,
change the it once the wear or damage is found
(
Or replace
every 3,000 lenses
)
.
>After the lens feeler is replaced, the lens feeler position needs to be recalibrated.
The calibration probe procedure is as follows
:
>>
>>
>>
>>
>>
>>
>>
>>
>>
>>
Dismountable lens feeler
Remove the screws and replace the lens feeler
#1 Enter the menu
#2
Enter
setup
interface
#3
Enter
setup
interface
#4 Take out tool
#5 Input correction tool
thickness value
#6 Replace the 25mm clamp
#7 Load calibration tool
#8 Start calibrating
#9 The machine starts the calibration procedure, saves the calibration parameters automatically, and
then takes out the calibration tool when the calibration is completed.
25mm
clamp