Available settings are: 50, 100, 250, 500, 1000, 1250, 1500, or 2000 microns. 
Factory default setting is “2000” microns. 
The default values are listed in microns, however values will be displayed in the current units selected. For 

example, if the user defined Unit Selection setting is Pascals, units displayed will be: 7, 14, 33, 66, 133, 166, 

200, and 266 (The Pascal conversion of the 50 – 2000 micron values).
The “ALARM”, “HIGH” icon will be displayed to indicate you are setting up the High Limit Alarm.
To scroll the selection, use the SCROLL UP and SCROLL DOWN buttons. 
To cancel, press the POWER button.
To select and store the setting shown, press the MODE button. When a setting is stored, you will be returned to 

main menu selections. Press POWER to return to normal operation mode.

High Limit Alarm

MODE and SCROLL UP buttons will enable the High Limit Alarm ONLY IF the existing pressure is below the 

preset, user defined High Alarm setting. If the existing pressure is higher than the preset, user defined High 

Alarm Limit, the High Limit Alarm cannot be enabled. If the Low Limit Alarm is currently enabled, setting the 

High Limit Alarm will disable the Low Limit Alarm even if the High Limit Alarm fails due to pressure readings 

above the High Limit Alarm setting. An enabled High Limit Alarm will display the “HIGH”, “ALARM” icon on 

the main display screen, and will disable the Low Alarm setting.
If High Limit Alarm is enabled and the pressure is greater than or equal to the preset, user defined High Alarm 

setting limit value, the alarm will alert with a beep sound every second and the “ALARM”, “HIGH” icon will 

flash. Beeping and flashing will continue even if the pressure drops below the High Limit Alarm setting. Press 

any button to turn off the alarm.

Button-press Feedback (BEEP)  

Key-press feedback determines whether an audible sound is heard each time a button is pressed. 
Available settings are: ON, OFF. Factory default setting is “ON”. 
To change the selection, use the SCROLL UP or SCROLL DOWN button. To cancel, press the POWER button.
To select and store the setting shown, press the MODE button. When a setting is stored, you will be returned to 

main menu selections. Press POWER to return to normal operation mode.

Display Resolution (HI RES) 

The display resolution menu allows you to select whether pressure will be shown at high resolution. High 

resolution means that pressure is shown as-is, in one unit increments. 
Available settings are: ON, OFF. Factory default setting is “OFF”. 
When the setting is OFF, the reading display is rounded.
To change the selection, use the SCROLL UP or SCROLL DOWN button. 
To cancel, press the POWER button.
To select and store the setting shown, press the MODE button. When a setting is stored, you will be returned to 

main menu selections. Press POWER to return to normal operation mode.
Note: Even if high resolution display is ON, pressure values above 1000 microns will be rounded off according 

to the current display pressure.

Summary of Contents for DV-41

Page 1: ...OPERATINGMANUAL DV 41 DIGITAL VACUUM GAUGE SUPERNOVA ...

Page 2: ...Supply AC adapter included or three AA alkaline batteries Dimensions 13 length x 11 width x 3 5 depth Shipping Weight 2 7 lbs 1 2 kg Warranty 12 months FEATURES Power Button Mode Button Battery Status Indicator Display Screen Hook Scroll Buttons Battery Cover back 1 4 SAE Flare Brass Fitting with Flexible 5 ft Cord Unit Selection M ...

Page 3: ...nStartUp Press POWER button once to power gauge on If unit display does not light up check for correct battery installation Startup display will show JB JUST BETTER a loading menu followed by either the micron value or the over range indicator oooooo The center bars in the loading menu move to indicate the unit is in initial warm up state It may take up to 30 seconds for the unit to go into normal...

Page 4: ...ng this time shutdown is canceled MENUSETTINGS AccessingMainMenuandSubMenuSettings 1 Press and hold MODE button for four seconds to access main menu settings 2 Press SCROLL UP and SCROLL DOWN buttons to cycle through all available main menu options UNIT AL LO AL HI BEEP HI RES AU OFF B LITE VER ID 3 To select a menu option press the MODE button to access sub menu Press the SCROLL UP and SCROLL DOW...

Page 5: ...ng is Pascals units displayed will be 7 14 33 66 133 166 200 and 266 The Pascal conversion of the 50 2000 micron values The ALARM LOW icon will be displayed to indicate you are setting up the Low Limit Alarm To scroll the selection use the SCROLL UP and SCROLL DOWN buttons To cancel press the POWER button To select and store the setting shown press the MODE button When a setting is stored you will...

Page 6: ... screen and will disable the Low Alarm setting If High Limit Alarm is enabled and the pressure is greater than or equal to the preset user defined High Alarm setting limit value the alarm will alert with a beep sound every second and the ALARM HIGH icon will flash Beeping and flashing will continue even if the pressure drops below the High Limit Alarm setting Press any button to turn off the alarm...

Page 7: ...settings are OFF 10 15 30 45 60 or 90 seconds Factory default setting is 10 To change the selection use the SCROLL UP and SCROLL DOWN buttons To cancel press the POWER button To select and store the setting shown press the MODE button When a setting is stored you will be returned to main menu selections Press POWER to return to normal operation mode Note While in the menu system the back light wil...

Page 8: ...r 12 months from date of purchase JB products are guaranteed when used in accordance with our guidelines and recommendations Warranty is limited to the repair replacement or credit at invoice price our option of products which in our opinion are defective due to workmanship and or materials In no case will we allow charges for labor expense or consequential damage Repairs performed on items out of...

Reviews: