3.1 Turning on device
a. Plug the power cable into a power outlet (Fig.
3.3) of the correct voltage and frequency as
defined on the manufacturer's technical label (Fig
2.2).
b. Press the power switch
(I/
O
)
(item 1 in Fig.2.1)
to the ON position
(I)
. The green indicator light
flashes until concentration is achieved.
IMPORTAINT!
TURN THE ANESTHESIA MACHINES
OXYGEN FLOW METER ON TO A
MINUMUM FLOW RATE OF 0.25 LPM
A “NO FLOW” ALARM MAY SOUND IF
THE FLOW METER ISN’T OPEN
3.2 Turning off device
At the end of the usage, press the
I/
O
Switch to
place it in the
O
(OFF) position to stop the device.
The oxygen enriched air flow continues for ap
-
proximately one minute after the device is stopped.
1. Filter / Silencer
2. Cabinet filter
3. Ventilation grill
Note: Shown with grate removed
Fig. 4.1
4. CLEANING - MAINTENANCE
Only the outside of the concentrator
is to be
cleaned. Use a damp sponge or cloth with
water only.
Acetone, solvents or any other inflammable
products
must not be used
.
Do not use abrasive powders.
FILTERS - IMPORTAINT!
Fine Filter (#1 fig. 4.1)
and Cabinet Filter Cleaning (#2)
The cabinet filter (#2 Fig.4.1) must be cleaned
in warm water and household detergent weekly
or after approximately
100 hours
of use. Dry
before reinstalling. More frequent cleaning is
recommended in dusty environments.
Cabinet filter: p/n OC4000-1
The fine intake filter should be replaced annually
or every 2000 hours of use. More frequently
dusty /dirty environments.
Fine filter p/n OC4000-2