PREVENTIVE MAINTENANCE

a. Wash cabinet filter weekly.
b. The fine intake filter should be replaced 

annually or every 2000 hours of use. More 

frequently dusty /dirty environments.

  

c. Check oxygen concentration every 15,000 

    hours or 3 years of operation to verify the

    continuing OCSI function.

Use original parts only

Contact Supera Anesthesia Innovations 

for additiional information

877-620-1500

 

Electrical power supply:

 

Rating: 

115/230V 60Hz     230V 50Hz                        

Average Power:     330W(avg)    

  300 W(avg) 

Protection Class:   Class II 

  Class II

Mains Protection:  5A  

  5A

Filters:

At the rear of the device: a cabinet air filter.

At the compressor input: an inlet air filter, 5 µm,  

located behind the cabinet air filter.   

Air circulation

 

A tubeaxial fan cools the compressor compart

-

ment.

Environmental limit conditions

The performances of the device (especially the 

oxygen concentration) are quoted at 21

o

C (70

o

F) 

and  one  atmosphere.  They  may  change  with 

temperature and altitude.
 • The device must be stored, transported and 

    used in the vertical position only.   

 •  Ambient temperature of between 5

o

C and

    40

o

C (40

o

to 104

o

F) operation.

 •  Storage temperature from -20

o

C to 60

o

    (-4

o

to 140

o

F).

 • Relative humidity of between 15% and 95%

    operation and storage, both non-condensing.  

 • Altitude(21

o

C): Up to 2,286m (7,500ft) 

    without degradation; Consult your equip-

    ment provider for further information re-

    garding 2,286m to 4000m (7500 to 13000 

    ft) 

 • Complies with EN60601-1 standard; spil-

    ling a glass of water.

5.7. Standards

EN 60601-1[UL60601-1:2003],

CAN/CSA-C22.2 No.601.1-M90 w/A1&A2: 

Electrical Safety- Medical Devices.
EN60601-1-2:2001  

Electromagnetic Compatibility

   

 

Summary of Contents for Pureline OC4000

Page 1: ...OC6000 Operators Manual Supera LLC Copyright 2019 OXYGEN CONCENTRATORS For Veterinary Use Only...

Page 2: ......

Page 3: ...away for help Our goal is to make your new machine as easy to use and care for as possible This device is meant to be operated under the normal surveillance and control of a veterinarian trained in it...

Page 4: ......

Page 5: ...th the 3 16 allen wrench provided VERY tightly 1 Carefully unpack the top assembly and frame from the box 2 Align the top to the frame as shown NOTE The frame base has two cross bars with holes in the...

Page 6: ...1 Remove the oxygen concentrator from the box and guide the 4 threaded post into the frame 2 Install and tighten the nuts DO NOT OVER TIGHTEN THE NUTS SNUG FIT IS ALL THAT IS NEEDED...

Page 7: ...Attach the green hose from the OC6000 top to the threaded fitting DO NOT OVER TIGHTEN Oxygen Setup Attach the green hose from your anesthesia machine to the DISS fitting...

Page 8: ...ch in the AUX position the supply of oxygen to the anesthesia machine comes from E tank on the back of the machine or another source using the E tank adapter p n OC6050 To use the E tank as a souce fo...

Page 9: ...F power switched off Class II protection Do not expose to open flames Do not use oil or grease Technical information Consult the accompanying documents Keep in the vertical position Fragile handle wit...

Page 10: ...manual should be allowed to operate the O2 concentrator TheCAUTIONSbelowindicateapotentially hazardous situation If conditions are not avoided a situation could occur that results in property damage o...

Page 11: ...Concentrator is packaged to protect the device from damage while being transported and stored Check for damage to the packaging After device is removed from the package inspect for damage If damage i...

Page 12: ...2 1 Front panel Fig 2 1 1 I O ON OFF switch 2 Indicator lights 3 Oxygen product outlet 4 Circuit breaker 5 Pre set flow meter non adjustable...

Page 13: ...s for ap proximatelyoneminuteafterthedeviceisstopped 1 Filter Silencer 2 Cabinet filter 3 Ventilation grill Note Shown with grate removed Fig 4 1 4 CLEANING MAINTENANCE Only the outside of the concent...

Page 14: ...he column in production The columns contain a molecular sieve whose function is to adsorbthenitrogenandthusallowoxygentopass The oxygen enriched product is then directed through a pressure reducing va...

Page 15: ...is fitted on the compressor outlet and is calibrated to 2 7 bar 40 psig 5 4 3 Indicators The green indicator light Fig 5 1 indicates that power is applied to the device When first turned on the indica...

Page 16: ...he maximum flow does not exceed 10 of the indicated value when a back pressure of 7 kPa 1 psig is applied to the output of the device The maximum outlet pressure is 50 kPa 7 psig 5 5 4 Maintenance of...

Page 17: ...ubeaxial fan cools the compressor compart ment Environmental limit conditions The performances of the device especially the oxygen concentration are quoted at 21o C 70o F and one atmosphere They may c...

Page 18: ...Pneumatic connection broken or other pressure problem Internal electrical fault Pneumatic circuit fault Dirty Filters blockage Fan is not working Observations The I O ON OFF button is in the I ON pos...

Page 19: ......

Page 20: ......

Page 21: ...machine This can be washed with mild soap and wa ter DO NOT INSTALL UNTIL 100 DRY P N 8400 1180 Fine inlet filter square black plastic filter List Price 14 95 The fine particle filter goes under the...

Page 22: ......

Page 23: ...al applicable for the product 3 products considered to be of a consumable nature such as rubber or plastic goods 4 accessories or parts not manufactured by Supera Anesthesia Innovations 5 specially ma...

Page 24: ...company s terminal where they are entirely responsible 5 Open and inspect your product as soon as possible DO NOT WAIT 6 If you find damage take as many photographs of everything as soon as you can a...

Page 25: ...DOCUMENTATION Date purchased Purchased from Machine serial number Machine service information Service Date Service information Service Date Service information Service Date Service information...

Page 26: ......

Page 27: ......

Page 28: ...OXYGEN CONCENTRATORS 503 723 5068 or 877 620 1500 toll free E Mail info superavet com 12330 S E Highway 212 Clackamas OR 97015 www SuperaVet com...

Reviews: