3.1 Turning on device

a.  Plug  the  power  cable  into  a  power  outlet 

(Fig. 3.3) of the correct voltage and frequency 

as defined on the manufacturer's technical label 

  

b. Press the power switch 

(I/

O

(item 1 in Fig.2.1)

 

to the ON position

 (I)

.  The green indicator light 

flashes until concentration is achieved.

TURN THE ANESTHESIA MACHINES OXYGEN 

FLOW METER ON TO A MINUMUM FLOW RATE 

OF 0.25 LPM

A "NO FLOW" ALARM WILL SOUND 

IF THE ANESTHESIA MACHINE  FLOW METER 

ISN'T OPEN

 

 

             

 

               

1. Filter / Silencer

           2. Cabinet filter

           3. Ventilation grill 

Note: Shown with grate removed

4. CLEANING - MAINTENANCE

Only the outside of the concentrator

 

is to be 

cleaned.    Use  a  damp  sponge  or  cloth  with  

water only.  
Acetone,  solvents  or  any  other  inflammable 

products 

must not be used

.  Do not use abra

-

sive powders.

Cabinet  air  filter

  (Item  1  Fig.4.1)  must  be 

cleaned in warm water and household detergent 

weekly or after approximately 100 hours of use. 

Dry before reinstalling. More frequent cleaning 

is recommended in dusty environments.

Fine filter

 (Item 2 Fig.4.2) remove the right 

side filter grate (Item 3 Fig.4.2) The fine intake 

filter  should  be  replaced  annually  or  every 

2000  hours  of  use.  More  frequently  dusty  /

dirty environments.

P/N OC8000-2

Fig. 4.1

1

2

3

Fig. 4.2

3.2 Turning off device

At the end of the usage, press the 

I/

O

 Switch  

to place it in the 

O

 

(OFF) position to stop the 

device. The oxygen enriched air flow continues 

for approximately one minute after the device 

is stopped.

Note:

 After turning the unit off, the user must 

wait  3-5  minutes  before  turning  it  back  on. 

System pressure must dissipate before the unit 

will properly restart.

FILTER MAINTENANCE

Summary of Contents for Pureline M8000

Page 1: ...Operators Manual Model OC8000 OC8200 M8000 not shown Supera LLC Copyright 2020 OXYGEN CONCENTRATORS For Veterinary Use Only...

Page 2: ......

Page 3: ...away for help Our goal is to make your new machine as easy to use and care for as possible This device is meant to be operated under the normal surveillance and control of a veterinarian trained in it...

Page 4: ......

Page 5: ...s with the 3 16 allen wrench provided VERY tightly 1 Carefully unpack the top assembly and frame from the box 2 Align the top to the frame as shown NOTE The frame base has two cross bars with holes in...

Page 6: ...1 Remove the oxygen concentrator from the box and guide the 4 threaded post into the frame 2 Install and tighten the nuts DO NOT OVER TIGHTEN THE NUTS SNUG FIT IS ALL THAT IS NEEDED...

Page 7: ...anesthesia machine to one of the threaded fittings NOTE Both output the same DO NOT OVER TIGHTEN Oxygen Hose Setup Attach the green hose from the OC8000 top assembly to the threaded fitting on the co...

Page 8: ...ch in the AUX position the supply of oxygen to the anesthesia machine comes from E tank on the back of the machine or another source using the E tank adapter p n OC6050 To use the E tank as a souce fo...

Page 9: ...tched on OFF power switched off Class II protection Do not expose to open flames Do not use oil or grease Technical information Consult the accompanying documents Keep in the vertical position Fragile...

Page 10: ...manual should be allowed to operate the O2 concentrator TheCAUTIONSbelowindicateapotentially hazardous situation If conditions are not avoided a situation could occur that results in property damage o...

Page 11: ...Concentrator is packaged to protect the device from damage while being transported and stored Check for damage to the packaging After device is removed from the package inspect for damage If damage i...

Page 12: ...2 1 Front panel Fig 2 1 1 I O ON OFF switch 2 Indicator lights 3 Oxygen product outlet 4 Circuit breaker 5 Flow meter...

Page 13: ...lammable products must not be used Do not use abra sive powders Cabinet air filter Item 1 Fig 4 1 must be cleanedinwarmwaterandhouseholddetergent weeklyorafterapproximately100hoursofuse Drybeforereins...

Page 14: ...ressor sends filtered room air to a solenoid valve which allows compressed air to pass to the column in production The columns contain a molecular sieve whose function is to adsorbthenitrogenandthusal...

Page 15: ...is fitted on the compressor outlet and is calibrated to 2 7 bar 40 psig 5 4 3 Indicators The green indicator light Fig 5 1 indicates that power is applied to the device When first turned on the indica...

Page 16: ...ssure Minimum recommended flow 2 lpm Maximum recommended flow 8 lpm The variation of the maximum recommended flow does not exceed 10 of the indicated value when a back pressure of 6 9 kPa 1 psig is ap...

Page 17: ...ncoolsthecompressorcompart ment and a second fan cools the heat exchanger coil Environmental limit conditions The performances of the device especially the oxygen concentration are quoted at 21o C 70o...

Page 18: ...ator Pneumatic connection broken or other pressure problem Internal electrical fault Pneumatic circuit fault Dirty Filters blockage Fan is not working Observations The I O ON OFF button is in the I ON...

Page 19: ......

Page 20: ......

Page 21: ...the machine This can be washed with mild soap and wa ter DO NOT INSTALL UNTIL 100 DRY P N OC8000 2 Fine inlet filter square black plastic filter The fine particle filter goes under the course filter T...

Page 22: ......

Page 23: ...al applicable for the product 3 products considered to be of a consumable nature such as rubber or plastic goods 4 accessories or parts not manufactured by Supera Anesthesia Innovations 5 specially ma...

Page 24: ......

Page 25: ...DOCUMENTATION Date purchased Purchased from Machine serial number Machine service information Service Date Service information Service Date Service information Service Date Service information...

Page 26: ...company s terminal where they are entirely responsible 5 Open and inspect your product as soon as possible DO NOT WAIT 6 If you find damage take as many photographs of everything as soon as you can a...

Page 27: ......

Page 28: ...503 723 5068 or 877 620 1500 toll free E Mail info superavet com 12330 S E Highway 212 Clackamas OR 97015 www SuperaVet com...

Reviews: